Programming Guide

9.5 Command List

In depending from the specific commands, the following characters are used:

?... host (e.g. PC) requests for a parameter value of the unit

!... unit acknowledges a valid parameter request and responses with the parameter value

=... host forces the unit to set a certain parameter

#... unit informs the host, a parameter was set on the control panel manually

*... unit’s error response

Description

Char

Format (2)

P (1)

B (1)

S (1)

N (1)

Legal Values

Factory Default

Burst string format

$

(3)

 

 

(3)

UTSI

Ambient radiation correction

A

nnnn

 

 

0000-3000°C

0000

(FA only)

 

 

 

 

 

 

 

 

Measured attenuation

B

nn

 

 

0 to 99

 

Advanced Hold Threshold

C

nnnn

 

 

0000-3000°C

0000 = no advanced hold

Baud rate (5)

D

nnn

 

 

 

003=300 baud

38400

 

 

 

 

 

 

 

012=1200 baud

 

 

 

 

 

 

 

 

024=2400 baud

 

 

 

 

 

 

 

 

096=9600 baud

 

 

 

 

 

 

 

 

192=19200 baud

 

 

 

 

 

 

 

384=38400 baud

 

Emissivity

E

n.nn

 

0.10-1.00

1.00

Valley hold time (FA only)

F

nnn.n

 

 

 

Average time (4)

G

nnn.n

 

000.0-300.0 s

000.0

Top of mA range

H

nnnn

0000-9999 (°C/°F)

High end of sensor range

Sensor internal ambient

I

nnn

 

 

 

 

Switch panel lock

J

X

 

 

L = locked, U = unlocked

unlocked

Relay alarm output control

K

n

 

 

 

0 = off

2

 

 

 

 

 

 

 

1 = on

 

 

 

 

 

 

 

 

2 = normally open

 

 

 

 

 

 

 

 

3 = normally closed

 

Bottom of mA range

L

nnnn

 

0000-9999 (°C/°F)

(6)

Mode (FR only)

M

n

1 = 1-color, 2 = 2-color

2

Target temperature

N

nnnn

 

 

 

1-color (FR only)

 

 

 

 

 

 

 

 

Output current

O

nn

 

 

00 = controlled by unit

00

 

 

 

 

 

 

 

02 = under range

 

 

 

 

 

 

 

 

21 = over range

 

 

 

 

 

 

 

 

00-20 = current in mA

 

Peak hold time (4)

P

nnn.n

000.0-300.0 s

0000.0

Table 5: Command List

(1)P = Poll Mode (Request for a parameters), B = Burst Mode (continuous sending of parameters in the burst string), S = Set (Command for setting a parameters), N = Notification (Acknowledgment for setting a parameter)

(2)n = number, X = uppercase letter

(3)see section 9.3.2 Burst Mode, p. 40

(4)Setting peak hold cancels average, and vice‐versa. 300.0 means reset only with external trigger

(5)The sensor restarts after a baud rate change.