Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–89
Equipment
required OIG501 (use with P6701B) (Item 27)
OIG502 (use with P6703B) (Item 28)
Optical Attenuator OA5022 (Item 29)
Fibre Optic Cable (Item 30)
Optical-to-Electrical Converter (item 31)
Prerequisites See page 1–15. Also, the probe and the oscilloscope channel it is
attached to must have been calibrated as a reference receiver.
1. Install the test hookup and preset the instrument controls:
Digitizing Oscilloscope
Optical to electrical converter
Output
Input
Optical Impulse
Generator
Output
Optical
Attenuator
Fiber optic cable
Figure 1–43: Reference-receiver performance-verification set up
a. Initialize the oscilloscope:
HPress SAVE/RECALL SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
b. Connect the probe, oscilloscope, optical impulse generator (OIG), and
the optical attenuator:
HConnect the optical probe to CH 1 of the oscilloscope under test (see
Figure 1–43).
HSee Table 1–9. If using a P6701B probe, connect the OIG501
(Item 27) OPTICAL OUTPUT to the optical attenuator (Item 29)
OPTICAL INPUT using a fiber-optic cable (Item 30). If using a
P6703B probe, connect the OIG502 (Item 28) OPTICAL OUTPUT
to the optical attenuator (Item 29) OPTICAL INPUT using a
fiber-optic cable (Item 30).
Reference ReceiverVerification