Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–91
HPress HORIZONTAL MENU; then press the main-menu button
Trigger Position. Set the Trigger Position to 20% using the general
purpose knob or keypad.
NOTE. Press SHIFT to change the horizontal position speed. The position moves
faster when SHIFT is lighted.
HAdjust the HORIZONTAL POSITION to move the 20% trigger
point onto the display (see Figure 1–44).
Figure 1–44: Optical impulse of Ch1 input from OA5022 Optical Attenuator
(OIG501 / OIG502 fed into optical attenuator in Step 1)
j. Select linear interpolation:
HIf SHIFT is lighted, press SHIFT.
HPress DISPLAY; then press the main-menu button Filter.
HPress the side-menu button Linear Interpolation.
2. Select the desired Mask standard and turn it on:
HPress MEASURE; then select Masks from the Measure pop-up menu.