Performance Tests
1–60 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Check output
Figure 1–21: Measurement of main trigger out limits
d. Check Delayed Trigger output against limits: See Figure 1–21.
HMove the precision 50 W cable from the rear-panel Main Trigger
Output BNC to the rear-panel Delayed Trigger Output BNC.
HCHECK that the Ch2 High readout is 1.0 volt and that the Ch2
Low readout 0.25 volts.
HEnter high and low voltages on test record.
HPress the side-menu button to select the 1 MW setting.
HPress CLEAR MENU.
HCHECK that the Ch2 High readout is 2.5 volts and that the Ch2
Low readout is 0.7 volts.
HEnter high and low voltages on test record.
3. Confirm CH 3 (Ax1 on some TDS models) output is within limits for gain:
a. Measure gain:
HMove the precision 50 W cable from the rear-panel DELAYED
TRIGGER OUTPUT BNC to the rear-panel SIGNAL OUT BNC.
HPush TRIGGER MENU.