Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–59
d. Modify the initialized front-panel control settings:
HSet the horizontal SCALE to 200 ms.
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Average.
HSelect 64 averages. Do this with the keypad or the general purpose
knob
2. Confirm Main and Delayed Trigger outputs are within limits for logic levels:
a. Display the test signal:
HPress WAVEFORM OFF to turn off CH 1.
HPress CH 2 to display that channel.
HSet the vertical SCALE to 1 V.
HUse the vertical POSITION knob to center the display on screen.
b. Measure logic levels:
HPress MEASURE; then press the main-menu button Select
Measurement for Ch2.
HSelect high and low measurements. To do this, repeatedly press the
side-menu button –more– until High and Low appear in the side
menu (their icons are shown at the left). Press both side-menu
buttons High and Low.
c. Check Main Trigger output against limits:
HCHECK that the Ch2 High readout is ≥2.5 volts and that the Ch2
Low readout is ≤0.7 volts. See Figure 1–21.
HEnter high and low voltages on test record.
HPress VERTICAL MENU; then press the main-menu button
Coupling. Now press the side-menu button to toggle it to the
50 W setting.
HCHECK that the Ch2 High readout is ≥1.0 volt and that the Ch2
Low readout ≤0.25 volts.
HEnter high and low voltages on test record.