Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–53
Equipment
required One sine wave generator (Item 19 or, optionally, items 25 and 26)
Two precision 50 coaxial cables (Item 5)
One 10X attenuator (Item 1)
One BNC T connector (Item 7)
One 5X attenuator (Item 2)
Prerequisites See page 1–15.
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
b. Modify the initialized front-panel control settings:
HSet the horizontal SCALE for the M (main) time base to 20 ns on
the TDS 600B or 25 ns on the TDS 500C/700C.
HPress HORIZONTAL MENU; then press the main-menu button
Time Base.
HPress the side-menu button Delayed Only; then press the side-menu
button Delayed Triggerable.
HSet the horizontal SCALE for the D (delayed) time base to 20 ns on
the TDS 600B or 25 ns on the TDS 500C/700C; then press the
side-menu button Main Only.
HPress TRIGGER MENU; then press the main-menu button Mode
& Holdoff. Press the side-menu button Normal.
HPress VERTICAL MENU; then press the main-menu button
Coupling. Press the side-menu button to select the 50 W setting.
HPress SHIFT; then press ACQUIRE MENU. Press the main-menu
button Mode; then press the side-menu button Average 16.
c. Hook up the test-signal source:
HConnect the signal output of a medium-frequency sine wave
generator (item 19 or, optionally, item 25) to a BNC T connector.
Connect one output of the T connector to CH 1 through a 50 W
precision coaxial cable. Connect the other output of the T connector
to the AUX TRIG INPUT at the rear panel. See Figure 1–18.
Sensitivity, Edge Trigger,DC Coupled