Brief Procedures
1–14 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
HTurn the general purpose knob to select the file to save. Choose
(or fd0:, the floppy disk drive, or hd0:, the optional
hard disk drive). With this choice, you will save a file starting with
, then containing 5-numbers, and a extension. For example,
the first time you run this on a blank, formatted disk or on the
Example Programs Disk, the oscilloscope will assign the name
to your file. If you ran the procedure again, the
oscilloscope would increment the name and call the file
.
f. To test the optional hard disk drive, use the general purpose knob to
select the file to save. Then choose the hard disk drive (hd0:). Save the
file as in step e.
HPress the side-menu button Save To Selected File.
2. Verify the file system works:
HPress the main-menu button Recall Factory Setup and the side-menu
button OK Confirm Factory Init to restore the 500 s time base and
the five cycle waveform.
HPress the main-menu button Recall Saved Setup; then press the
side-menu button From File.
HTurn the general purpose knob to select the file to recall. For example, if
you followed the instructions previously and saved the file to either the
floppy disk drive (fd0:), or the optional hard disk drive (hd0:), you had
the oscilloscope assign the name TEK00000.SET to your file.
HPress the side-menu button Recall From Selected File.
HVerify that digitizing oscilloscope retrieved the saved setup from the
disk. Do this by noticing the horizontal SCALE for the M (main) time
base is again 200 s and the waveform shows only two cycles just as it
was when you saved the setup.
3. Remove the test hookup:
HDisconnect the probe from the channel input and the probe-compensa-
tion terminals.
HRemove the floppy disk from the floppy disk drive.