Performance Tests
1–66 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Figure 1–25: Measurement of probe compensator amplitude
g. Check against limits:
HSubtract the value just obtained (base level) from that obtained
previously (top level).
HCHECK that the difference obtained is within 495 mV to 505 mV,
inclusive.
HEnter voltage difference on test record.
3. Disconnect the hookup: Disconnect the cable from CH 1.