Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–63
c. Modify the initialized front-panel control settings:
HSet the horizontal SCALE to 200 s.
HPress SET LEVEL TO 50%.
HUse the vertical POSITION knob to center the display on screen.
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Average.
HSelect 128 averages with the keypad or the general purpose knob.
2. Confirm that the Probe Compensator signal is within limits for frequency:
a. Measure the frequency of the probe compensation signal:
HPress MEASURE; then press the main-menu button Select
Measrmnt for Ch1.
HRepeatedly press the side-menu button –more– until Frequency
appears in the side menu (its icon is shown at the left). Press the
side-menu button Frequency.
b. Check against limits:
HCHECK that the CH 1 Freq readout is within 950 Hz to 1.050 kHz,
inclusive. See Figure 1–23.
HEnter frequency on test record.
HPress MEASURE; then press the main-menu button Remove
Measrmnt. Press the side-menu Measurement 1.