Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–37
Set the generator
(reference) frequency to
the test frequency from
Table 1–4.
1
Set the horizontal scale
from Table 1–4. 2
Read results. 3
Figure 1–9: Measurement of analog bandwidth
g. Check against limits:
HCHECK that the Pk-Pk readout on screen is within the limits listed
in Table 1–4 for the current vertical scale setting.
HEnter voltage on test record.
HWhen finished checking, set the horizontal SCALE back to the
50 ns setting.
STOP. Checking each channel’s bandwidth at all vertical scale settings is time
consuming and unnecessary. You may skip checking the remaining vertical scale
settings in Table 1–4 (that is, skip the following substep, h) if this digitizing
oscilloscope has performed as follows:
HPassed the 100 mV vertical scale setting just checked in this
procedure.
HPassed the Verify Internal Adjustment, Self Compensation, and
Diagnostics procedure found under Self Tests, on page 1–5.