Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–29
Align the active cursor
over the waveform. 1
Read the
measurement
results. 2
Figure 1–4: Measurement of DC offset accuracy at zero setting
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
WARNING. The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure.
Equipment
required Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision 50 W coaxial cables (Item 5)
Prerequisites The oscilloscope must meet the prerequisites listed on page 1–15
See Input Channels vs. Model on page 1–2.
Check DC VoltageMeasurement Accuracy