Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–77
PAL signal
source Digitizing Oscilloscope
75 cable
75 terminator
Output
Figure 1–35: Line count accuracy test hookup
HPress the main-menu Trigger Position.
HPress the side-menu to Set to 50%.
HPress the main-menu to Horiz Pos.
HPress the side-menu to Set to 50%.
HUse the HORIZONTAL POSITION knob to move the falling edge
of the sync pulse to two divisions to the left of center screen. See
Figure 1–36.