Performance Tests
1–18 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications Table 1–1: Test equipment (cont.)Item number and
description PurposeExampleMinimum requirements
27. Generator, Optical Im-
pulse 850 nm optical impulse Tektronix OIG501 Optical
Impulse Generator1Checking Option 3C
28. Generator, Optical Im-
pulse 1300 nm optical impulse Tektronix OIG502 Optical
Impulse Generator1Checking Option 4C
29. Attenuator, Optical 62.5 micron optical attenuator Tektronix OA5022 Optical
Attenuator1Checking Option 3C and 4C
30. Cable, Fibre Optic FC-FC fiber cable 174-2322-00 Checking Option 3C and 4C
31. Optical-to-Electrical
Converter P6701B (used with Option 3C) or
P6703B (used with Option 4C)
optical-to-electrical converter
Tektronix part number
P6701B or P6703B Checking Option 3C and 4C
1Requires a TM 500 or TM 5000 Series Power Module Mainframe.
2Warning: The P6243 and P6245 probes that may be used with this oscilloscope provide an extremely low loading
capacitance (<1 pF) to ensure the best possible signal reproduction. These probes should not be used to measure
signals exceeding ±8 V, or errors in signal measurement will be observed. Above 40 V, damage to the probe may result.
To make measurements beyond ±8 V, use either the P6139A probe (good to 500 V), or refer to the catalog for a recom-
mended probe.
3You can replace items 19, 20, or 21 with a Tegam/T ektronix SG503 (item 25) or SG504 (item 26) – if available.