Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications 1–75
HReconnect the output of the signal generator. Connect the composite
signal connector of the PAL signal source (labeled COMPST on the
TSG 121) to a 75 W cable and a 75 W terminator. Connect both
signals to the CH1 input through a BNC T. See Figure 1–33.
HPress VERTICAL MENU.
HIf needed, press the main-menu Fine Scale.
HUse the keypad to set fine scale to 500 mV (press 500, SHIFT, m,
then ENTER).
HConnect another composite signal connector of the PAL signal
source (labeled COMPST on the TSG 121) through a 75 W cable
and a 75 W terminator to the CH2 input. See Figure 1–33.
PAL signal
source
BNC T connector
75 terminators
Digitizing Oscilloscope
50 cable
Signal Generator
75 cable
75 Cable
COMPST
COMPST
Output
Figure 1–33: Subsequent 60 Hz Rejection test hookup
HCONFIRM that the TRIG’D LED stays lighted and that the
waveform on screen is stable. In other words, be sure the waveform
does not move horizontally or vertically. Also, confirm that the
waveform on the screen has one positive pulse and a number of
negative pulses. See Figure 1–34.
HDisconnect all test equipment from the digitizing oscilloscope.