Specifications2–4 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications Table 2–2: Record length and divisions per record vs. TDS model
S a ar el
Rec r le
Divisions per record
S
t
a
nd
ar
d Mod
el
s
Rec
o
r
d
le
ngth FTS1 Off2FTS1 On3
All
T
D
S 500C, T
D
S600B & T
D
S 700C
All ha els4
500 10 divs 10 divs
All
c
ha
nn
els4
1,000 20 divs 10 divs
2,500 50 divs 10 divs
5,000 100 divs 10 divs
15,000 300 divs 10 divs
TDS 500C & TDS 700C,
All channels 50,000 1,000 divs 10 divs
el i pi 1
Rec r le
Divisions per record
Mod
el
s w
i
th O
p
t
i
on
1
M
Rec
o
r
d
le
ngth FTS1 Off2FTS1 On3
TDS 500C & TDS 700C,
All channels 75,000 1,500 divs 10 divs
TDS 500C & TDS 700C,
All channels 100,000 2,000 divs 10 divs
TDS 500C & TDS 700C,
All channels 130,000 2,600 divs 10 divs
250,000 5,000 divs 10 divs
500,000 10,000 divs 10 divs
el i pi 2
Rec r le
Divisions per record
Mod
el
s w
i
th O
p
t
i
on
2
M
Rec
o
r
d
le
ngth FTS1 Off2FTS1 On3
TDS 520C & TDS 724C
One or two channels
TDS 540C, TDS 784C & TDS 754C
Three or four channels
2,000,000 40, 000 divs 10 divs
TDS 520C & TDS 724C
One channel only
TDS 540C, TDS 784C & TDS 754C
Two channels
4,000,000 80,000 divs 10 divs
TDS 540C, TDS 784C & TDS 754C
One channel only 8,000,000 160,000 divs 10 divs
1Fit to Screen setting
2Fit to Screen off preserves 50 samples/division in a 1–2–5 sec/division sequence.
3Fit to Screen on lets the samples/division and the sec/division sequence vary.
4All channels means all that may be displayed at one time: four channels for some
models, two for others. See Table 2–1 and its footnote on page 2–1.