Performance Tests
1–80 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
HPress the side-menu Fields.
HUse the keypad to set the number of fields to 1 (press 1, then
ENTER).
HPress the side-menu Sync Width.
HUse the keypad to set the width to 400 ns (press 400, SHIFT, n,
then ENTER).
HPress the side-menu –more– 1 of 2. Then press V1 Start Time.
HUse the keypad to set V1 start time to 10.10 s (press 10.10, SHIFT,
m, then ENTER).
HPress the side-menu V1 Stop Time.
HUse the keypad to set V1 stop time to 10.50 s (press 10.50, SHIFT,
m, then ENTER).
HPress the main-menu Type pop-up to select Edge.
HPress HORIZONTAL MENU.
HPress the main-menu Record Length.
HSelect the side-menu 1000 points in 20div. If needed, first press the
side-menu –more– until you see the appropriate side-menu item.
HTurn the HORIZONTAL POSITION knob to position the trigger
‘T’ two divisions to the left of the center screen.
HPress MEASURE.
HIf needed, press the main-menu Select Measrmnt.
HPress the side-menu Negative Width.
HPress the side-menu Period.
b. Set up the pulse generator for Sync Duty Cycle Test:
HSet PULSE DURATION to 50 ns.
HSet PERIOD to 10 s.
HSet OUTPUT (VOLTS) to –1 for LOW LEVEL and +1 for HIGH
LEVEL.
HDepress the COMPLEMENT button.
HBe sure BACK TERM is depressed (in).