Front Panel Operation - 5
Scrolling Keys
QR
QR
Step Step
Q
Channel
Scrolling keys let you move through the commands in the presently selected function menu. Press ▼ to bring up the next command in the list. Press ▲ to go back to the previous command in the list. Function menus are circular; you can return to the starting position by continuously pressing either key
These keys let you scroll through points in a list function. When the display indicates EOL, the end of the list has been reached.
Press this key to scroll through all channels in the mainframe
Metering Keys
Metering keys control the metering functions of the electronic load. As set from the factory, all front panel measurements are calculated from a total of 1000 readings taken at a 10 microsecond sampling rate. Therefore, the factory default acquisition time for a single front panel measurement is about 10 milliseconds. Refer to “Making Front Panel Measurements” for more information about changing the front panel sampling rate and the number of measurement points.
Meter
Takes the front panel back to the Metering mode from any other mode.
Also accesses the meter menu. This menu lets you measure various parameters of the selected input.
Display | Measurement |
<chan> <voltage> <current> | Displays the voltage and current of the selected channel |
<chan> <value> V MAX | Displays the maximum voltage |
<chan> <value> V MIN | Displays the minimum voltage |
<chan> <value> V RMS | Displays the rms voltage |
<chan> <value> A MAX | Displays the maximum current |
<chan> <value> A MIN | Displays the minimum current |
<chan> <value> A RMS | Displays the rms voltage |
<chan> <value> WATTS | Displays the input power |
<chan> <value> W MAX | Displays the maximum power |
<chan> <value> W MIN | Displays the minimum power |
Shift | Sense | Specifies the measurement functions and ranges. | |
|
| Display | Command Function |
|
| S:PNT | Defines the number of data points in the measurement |
|
| S:TIN | Sets the digitizer sample spacing |
|
| S:WIN | Sets the measurement window function |
|
| S:OFF | Defines the data offset in the measurement |
|
| S:C:RNG | Selects the current measurement range |
|
| S:V:RNG | Selects the voltage measurement range |
Notes:Use
Use
R
RInput
and and
Q
QInput
to scroll through the menu commands
to scroll through the command parameters