B
Performance Test and Calibration Procedures
Introduction
This appendix contains test procedures for checking the operation and calibration of the Agilent N330xA Series Electronic Load Modules. The tests are performed using the front panel keypad of the N3300A or N3301A Mainframes. The required test equipment is listed in Table
The calibration ( adjustment ) procedures create new calibration constants to bring the current, voltage and resistance values within specifications. Calibration cannot be performed using the front panel keypad, an IEEE- 488 ( GPIB ) controller is required. Note that the Electronic Load Mainframe does not require calibration.
Important Perform the verification procedures before calibrating your Electronic Load Module. If the Electronic Load Module passes the verification procedures, the unit is operating within its calibration limits and does not need to be
If the electronic load requires service, refer to the list of Agilent Sales and Support Offices at the back of this manual.
Equipment Required
The equipment listed in the following table, or the equivalent to this equipment, is required for verification and calibration. An
Table B-1. Equipment Required
Equipment | Characteristics | Recommended Models |
Current Shunts1 | 0.1 Ohms @ 15 Amps, 0.04% @ 25 watts | Guildline 9230/15 |
| 0.01 Ohms @ 100 Amps, 0.04% @ 100 watts | Guildline 9230/100 |
| 0.001 Ohms @ 300 Amps, 0.04% @ 100 watts | Guildline 9230/300 |
Voltmeter | DC accuracy 0.01% 5 digit | Agilent 34401A or 3458A |
Current Probe | DC to 100Arms - DC to 100kHz | Agilent 1146A |
Oscilloscope |
| Agilent 54520A |
Power Source | 61Vdc / 61A minimum ( N3302A, N3304A, N3306A ) | 6032A or equivalent |
| 20Vdc / 143A minimum ( N3305A, N3306A ) | 6031A or equivalent |
| 241Vdc / 10A minimum ( N3303A, N3305A ) | 6035A |
Controller | GPIB ( IEEE 488.2 ) | HP Series 200/300 or |
|
| PC with HP 82341 IEEE |
|
| 488.2 Interface card and |
|
| Basic for Windows or equiv. |
1A
75