Agilent Technologies n3301a manual Performance Test and Calibration Procedures, Equipment Required

Page 72

B

Performance Test and Calibration Procedures

Introduction

This appendix contains test procedures for checking the operation and calibration of the Agilent N330xA Series Electronic Load Modules. The tests are performed using the front panel keypad of the N3300A or N3301A Mainframes. The required test equipment is listed in Table B-1 and sample performance test record cards are included at the end of the performance test section. The performance tests confirm the Agilent N330xA Series Electronic Load Modules meet all their published specifications.

The calibration ( adjustment ) procedures create new calibration constants to bring the current, voltage and resistance values within specifications. Calibration cannot be performed using the front panel keypad, an IEEE- 488 ( GPIB ) controller is required. Note that the Electronic Load Mainframe does not require calibration.

Important Perform the verification procedures before calibrating your Electronic Load Module. If the Electronic Load Module passes the verification procedures, the unit is operating within its calibration limits and does not need to be re-calibrated.

If the electronic load requires service, refer to the list of Agilent Sales and Support Offices at the back of this manual.

Equipment Required

The equipment listed in the following table, or the equivalent to this equipment, is required for verification and calibration. An IEEE-488.2 controller is not required for verification.

Table B-1. Equipment Required

Equipment

Characteristics

Recommended Models

Current Shunts1

0.1 Ohms @ 15 Amps, 0.04% @ 25 watts

Guildline 9230/15

 

0.01 Ohms @ 100 Amps, 0.04% @ 100 watts

Guildline 9230/100

 

0.001 Ohms @ 300 Amps, 0.04% @ 100 watts

Guildline 9230/300

Voltmeter

DC accuracy 0.01% 5 digit

Agilent 34401A or 3458A

Current Probe

DC to 100Arms - DC to 100kHz

Agilent 1146A

Oscilloscope

 

Agilent 54520A

Power Source

61Vdc / 61A minimum ( N3302A, N3304A, N3306A )

6032A or equivalent

 

20Vdc / 143A minimum ( N3305A, N3306A )

6031A or equivalent

 

241Vdc / 10A minimum ( N3303A, N3305A )

6035A

Controller

GPIB ( IEEE 488.2 )

HP Series 200/300 or

 

 

PC with HP 82341 IEEE

 

 

488.2 Interface card and

 

 

Basic for Windows or equiv.

1A 4-terminal resistor ( current shunt ) is required to eliminate output current measurement errors caused by voltage drops in the load leads and connections.

75

Image 72
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceInstrument Configuration Rear Panel At a GlanceQuick Reference Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Task Where to find information Document OrientationGetting Started Map Safety Considerations Options and AccessoriesDescription Features and Capabilities Front Panel ControlsRemote Programming Constant Current CC Mode Operating ModesImmediate Current Level Constant Resistance CR Mode General Information Triggered Current LevelTransient Current Level Software Current LimitConstant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Triggered Voltage Level Transient Voltage LevelTransient Operation Continuous Transient OperationPulsed Transient Operation Toggled Transient OperationTriggering a preset level List OperationTriggered Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Protection Features Resetting Latched ProtectionInput Control Short On/OffOverpower OvertemperatureGeneral Information Overvoltage OvercurrentReading Remote Programming Errors Reverse VoltageStatus Reporting Saving and Recalling Settings External Control SignalsRemote Sensing Monitor OutputsFault Input MeasurementsPort On/Off Power Measurements DC MeasurementsRMS Measurements Minimum and Maximum MeasurementsCleaning InspectionDamage Packaging Material Items SuppliedInstalling the Modules ProcedureInstallation Location Channel Number3A. N3300A Outline Diagram Bench Operation Rack MountingDo not block the fan exhaust at the rear of the unit Power Cord Manually-Tightened ConnectorsInput Connections 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Stranded Copper Wire Ampere Capacity Wire Size AmpacityMaximum Wire Lengths to Limit Voltage Drops Control Connector A1-A3A11 and A10 +S and -SSense Switch Trigger and Digital ConnectionsComputer Connections Gpib InterfaceRS-232 Interface Application Connections Local Sense ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingCheckout Procedure IntroductionCheckout Programming Values Error Messages Selftest ErrorsSelftest Error Codes Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysScrolling Keys Metering KeysDisplay Measurement Input Control Keys InpshorTransient Control Keys Trigger Control KeysList Control Keys Entry Keys Entry KeysUsing the Front Panel Display Set the constant current mode inputExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Querying and Clearing Output Protection and Errors Query and Clear Errors as followsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsSetting the Gpib Address Storing and Recalling Instrument StatesSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredCC Mode Tests Performance TestsImon Zero Verification Performance Test and CalibrationCV Mode Tests CC Mode Regulation TestPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test CR Mode Tests Resistance Range 1 Programming AccuracyResistance Range 2 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Table B-2 Variables and Power Source Settings CalibrationParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 United States Latin America Canada Australia/New ZealandEurope Asia Pacific JapanManual Updates