B - Performance Test and Calibration

1690 CLEAR SCREEN

1700 !

1710 OUTPUT @Ld;"INP:STAT ON"

1720 OUTPUT @Ld;"CAL:LEV P1"

1730 !

1740 PRINT TABXY(10,10),"Calibrating low current range P1 - 30 SECOND WAIT" 1750 WAIT 30

1760 CLEAR SCREEN

1770 !

1780 INPUT "Measure voltage across current shunt and enter in volts",Ip1 1790 !

1800 Ip1=Ip1/Rshunt

1810 !

1820 OUTPUT @Ld;"CAL:DATA ";Ip1

1830 !

1840 OUTPUT @Ld;"CAL:LEV P2"

1850 !

1860 PRINT TABXY(10,10),"calibrating low current range P2 - 30 SECOND WAIT" 1870 WAIT 30

1880 CLEAR SCREEN

1890 !

1900 INPUT "Measure voltage across current shunt and enter in volts",Ip2 1910 !

1920 Ip2=Ip2/Rshunt

1930 !

1940 OUTPUT @Ld;"CAL:DATA ";Ip2

1950 !

1960 OUTPUT @Ld;"CAL:SAVE"

1970 OUTPUT @Ld;"CAL:STAT OFF"

1980 !

1990 PRINT "IMON, IPROG and Current calibration adjustments complete"

2000 PRINT "Verify Current Adjustments"

2010 !

2020 END

94

Image 91
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols Declaration of Conformity DeclarationPrinting History Acoustic Noise InformationTable of Contents TURN-ON Checkout 101 Front Panel -At a Glance Quick ReferenceRear Panel At a Glance Instrument ConfigurationQuick Reference Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance Voltage Quick Reference Front Panel MenusComplete description of all programming commands Scpi Programming Commands At a GlanceQuick Reference Scpi Commands Document Orientation Task Where to find informationGetting Started Map Options and Accessories Safety ConsiderationsDescription Front Panel Controls Features and CapabilitiesRemote Programming Operating Modes Constant Current CC ModeImmediate Current Level Software Current Limit Constant Resistance CR ModeGeneral Information Triggered Current Level Transient Current LevelTriggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Transient Voltage Level Triggered Voltage LevelToggled Transient Operation Transient OperationContinuous Transient Operation Pulsed Transient OperationTriggering a transient pulse Triggering a preset levelList Operation Triggered OperationRisetime Transition Limitation Slew Rate and Minimum Transition TimeTransition Time Slew Rate Examples Short On/Off Protection FeaturesResetting Latched Protection Input ControlOvercurrent OverpowerOvertemperature General Information OvervoltageReverse Voltage Reading Remote Programming ErrorsStatus Reporting Monitor Outputs Saving and Recalling SettingsExternal Control Signals Remote SensingInput Measurements FaultPort On/Off Minimum and Maximum Measurements Power MeasurementsDC Measurements RMS MeasurementsItems Supplied CleaningInspection Damage Packaging MaterialProcedure Installing the ModulesInstallation Channel Number Location3A. N3300A Outline Diagram Rack Mounting Bench OperationDo not block the fan exhaust at the rear of the unit Manually-Tightened Connectors Power CordInput Connections 6a. Manual Connector 8mm Screw Terminal Connector option UJ1Wire Considerations Wire Size Ampacity Stranded Copper Wire Ampere CapacityMaximum Wire Lengths to Limit Voltage Drops +S and -S Control ConnectorA1-A3 A11 and A10Trigger and Digital Connections Sense SwitchGpib Interface Computer ConnectionsRS-232 Interface Parallel Connections Application ConnectionsLocal Sense Connections Remote Sense Connections11. Remote Sensing 13. Zero Volt Loading Low Voltage OperationIntroduction Checkout ProcedureCheckout Programming Values Case of Trouble Error MessagesSelftest Errors Selftest Error CodesFront Panel, Overall View Front Panel DescriptionFront Panel Operation System Keys Display Command FunctionImmediate Action Keys Function keysMetering Keys Scrolling KeysDisplay Measurement Inpshor Input Control KeysTrigger Control Keys Transient Control KeysList Control Keys Entry Keys Entry KeysProgramming Constant Current, Voltage and Resistance Modes Using the Front Panel DisplaySet the constant current mode input Examples of Front Panel ProgrammingSet the constant voltage mode input Set the constant resistance mode inputProgramming Transient Operation Set transient operation for Continuous modeSet transient operation for Toggled mode Set transient operation for Pulse modeTmode Togg Programming ListsList Sequence Use the Meter menu for making front panel measurements Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows Making Basic Front Panel MeasurementsSave an instrument state in location 0 as follows Setting the Gpib AddressStoring and Recalling Instrument States Set the Gpib address as followsSpecifications Specifications Table A-2. Supplemental Characteristics Specifications aN3302A N3303A N3304A N3305A N3306A SpecificationsEquipment Required Performance Test and Calibration ProceduresPerformance Test and Calibration CC Mode TestsPerformance Tests Imon Zero VerificationVoltage Programming and Measurement Accuracy Test CV Mode TestsCC Mode Regulation Test Performance Test and Calibration BCV Mode Regulation Test Resistance Range 1 Programming Accuracy CR Mode TestsResistance Range 2 Programming Accuracy Resistance Range 4 Programming Accuracy Resistance Range 3 Programming AccuracyFrequency and Duty Cycle Accuracy Transient Generator Mode TestsOutput 705TRAN on Output 705TRIGIMM Performance Test and Calibration Pulse Width AccuracyAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Calibration Table B-2 Variables and Power Source SettingsParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Print TABXY9,15CALIBRATING High Voltage Range Voltage Calibration ProgramPerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 101 Index102 Index103 Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates