Agilent Technologies n3302a, n3305a, n3304a, n3301a, n3303a Agilent N3304A Verification Test Record

Page 82

Performance Test and Calibration - B

Agilent N3304A Verification Test Record

Test Description

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

60 Ampere Range Programming and Readback

0984 A

_______A

1.016 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout –10.5mA

_______A

Aout + 10.5mA

High Current ( 60 A )

 

59.925 A

_______A

60.075 A

Front Panel Display

 

Aout – 40mA

_______A

Aout +40mA

6 Ampere Range Programming and Readback

0.9915

_______A

1.0085

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 5.5mA

_______A

Aout + 5.5mA

High Current ( 6 A )

 

5.9865 A

_______A

6.0135 A

Front panel Display

 

Aout – 8mA

_______A

Aout + 8mA

 

 

 

 

 

 

Current Regulation Test

 

 

5 Amp Regulation at input voltage delta 4V to 60V

 

______mA

Less then 10 mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

60 Volt Range Programming and Readback

2.957 V

_______V

3.043 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 21mV

_______V

Vout + 21mV

High Voltage ( 60 V )

 

59.932 V

_______V

60.068 V

Front Panel Display

 

Vout – 140mV

_______V

Vout + 140mV

6 Volt Range Programming and Readback

2.9905 V

_______V

3.0095 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 4.5mV

_______V

Vout + 4.5mV

High Voltage ( 6 V )

 

5.991 V

_______V

6.009 V

Front Panel Display

 

Vout – 6mV

_______V

Vout + 6mV

 

 

 

 

 

 

Voltage Mode

Regulation Test

 

 

5V regulation with input current delta 1A to 60A

 

______mV

Less then 10mV

 

 

 

 

 

 

Resistance

Mode Tests

 

 

Resistance Range 1

 

 

______ Ω

2.020 Ω

High Resistance Point

2 Ω

1.980 Ω

Low Resistance Point

0.50 Ω

0.486 Ω

______ Ω

0.514 Ω

Resistance Range 2

 

1.733 Ω

______Ω

1.867 Ω

Low Resistance Point

1.8 Ω

High Resistance Point

20 Ω

19.360 Ω

______ Ω

20.640 Ω

Resistance Range 3

 

16.800 Ω

______Ω

19.20 Ω

Low Resistance Point

18 Ω

High Resistance Point

200 Ω

159.88 Ω

______Ω

240.12 Ω

Resistance Range 4

 

144 Ω

______Ω

298.8 Ω

Low Resistance Point

180 Ω

High Resistance Point

2000 Ω

1000 Ω

______Ω

42,000 Ω

 

 

 

 

 

 

Transient Generator Tests

 

 

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

 

85

Image 82
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceRear Panel At a Glance Instrument ConfigurationQuick Reference Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Document Orientation Task Where to find informationGetting Started Map Options and Accessories Safety ConsiderationsDescription Front Panel Controls Features and CapabilitiesRemote Programming Operating Modes Constant Current CC ModeImmediate Current Level Transient Current Level Constant Resistance CR ModeGeneral Information Triggered Current Level Software Current LimitTriggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Triggered Voltage Level Transient Voltage LevelPulsed Transient Operation Transient OperationContinuous Transient Operation Toggled Transient OperationTriggered Operation Triggering a preset levelList Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Input Control Protection FeaturesResetting Latched Protection Short On/OffGeneral Information Overvoltage OverpowerOvertemperature OvercurrentReverse Voltage Reading Remote Programming ErrorsStatus Reporting Remote Sensing Saving and Recalling SettingsExternal Control Signals Monitor OutputsInput Measurements FaultPort On/Off RMS Measurements Power MeasurementsDC Measurements Minimum and Maximum MeasurementsDamage Packaging Material CleaningInspection Items SuppliedProcedure Installing the ModulesInstallation Location Channel Number3A. N3300A Outline Diagram Rack Mounting Bench OperationDo not block the fan exhaust at the rear of the unit Manually-Tightened Connectors Power CordInput Connections 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Wire Size Ampacity Stranded Copper Wire Ampere CapacityMaximum Wire Lengths to Limit Voltage Drops A11 and A10 Control ConnectorA1-A3 +S and -SSense Switch Trigger and Digital ConnectionsGpib Interface Computer ConnectionsRS-232 Interface Remote Sense Connections Application ConnectionsLocal Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingIntroduction Checkout ProcedureCheckout Programming Values Selftest Error Codes Error MessagesSelftest Errors Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysMetering Keys Scrolling KeysDisplay Measurement Input Control Keys InpshorTrigger Control Keys Transient Control KeysList Control Keys Entry Keys Entry KeysExamples of Front Panel Programming Using the Front Panel DisplaySet the constant current mode input Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Making Basic Front Panel Measurements Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows Use the Meter menu for making front panel measurementsSet the Gpib address as follows Setting the Gpib AddressStoring and Recalling Instrument States Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredImon Zero Verification CC Mode TestsPerformance Tests Performance Test and CalibrationPerformance Test and Calibration B CV Mode TestsCC Mode Regulation Test Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test Resistance Range 1 Programming Accuracy CR Mode TestsResistance Range 2 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Calibration Table B-2 Variables and Power Source SettingsParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 Europe Asia Pacific United States Latin AmericaCanada Australia/New Zealand JapanManual Updates