Agilent Technologies n3303a, n3305a, n3304a, n3301a, n3306a Agilent N3302A Verification Test Record

Page 80

Performance Test and Calibration - B

Agilent N3302A Verification Test Record

Test Description

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

30 Ampere Range Programming and Readback

0.989 A

_______A

1.011 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 6.5mA

_______A

Aout + 6.5mA

High Current ( 30 A )

 

29.960 A

_______A

30.040 A

Front Panel Display

 

Aout – 21mA

_______A

Aout + 21mA

3 Ampere Range Programming and Readback

0.994 A

_______A

1.006 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 3.5mA

_______A

Aout + 3.5mA

High Current ( 3 A )

 

2.992 A

_______A

3.008 A

Front panel Display

 

Aout – 4.5mA

_______A

Aout + 4.5mA

 

Constant Current

Regulation Test

 

 

2.5 Amp Regulation at input voltage delta 4V to 60V

 

______ mA

Less then 10mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

60 Volt Range Programming and Readback

2.989 V

_______V

3.011 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 9.5mV

_______V

Vout + 9.5mV

High Voltage ( 60 V )

 

59.932 V

_______V

60.068 V

Front Panel Display

 

Vout – 3.8mV

_______V

Vout + 3.8mV

6 Volt Range Programming and Readback

2.9905 V

_______V

3.0095 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 4.5mV

_______V

Vout + 4.5mV

High Voltage ( 6 V )

 

5.991 V

_______V

6.009 V

Front Panel Display

 

Vout – 6.0mV

_______V

Vout + 6.0mV

 

 

 

 

 

 

Voltage Mode

Regulation Test

 

 

5V regulation with input current delta 1A to 30A

 

______mV

Less then 5mV

 

 

 

 

 

 

Resistance

Mode Tests

 

 

Resistance Range 1

 

3.944 Ω

______ Ω

4.056 Ω

High Resistance Point

4 Ω

Low Resistance Point

0.1 Ω

0.0756 Ω

______ Ω

0.1244 Ω

Resistance Range 2

 

 

 

 

Low Resistance Point

3.6 Ω

3.466 Ω

______Ω

3.734 Ω

High Resistance Point

40 Ω

38.520 Ω

______Ω

41.480 Ω

Resistance Range 3

 

 

 

 

Low Resistance Point

36 Ω

32.880 Ω

______Ω

39.120 Ω

High Resistance Point

400 Ω

271.76 Ω

______ Ω

528.24 Ω

Resistance Range 4

 

 

 

 

Low Resistance Point

360 Ω

270 Ω

______Ω

720 Ω

High Resistance Point

2000 Ω

1200 Ω

______Ω

10,000 Ω

 

 

 

 

 

 

Transient Generator Tests

 

 

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

 

83

Image 80
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceQuick Reference Instrument ConfigurationRear Panel At a Glance Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Getting Started Map Task Where to find informationDocument Orientation Description Safety ConsiderationsOptions and Accessories Remote Programming Features and CapabilitiesFront Panel Controls Immediate Current Level Constant Current CC ModeOperating Modes Constant Resistance CR Mode General Information Triggered Current LevelTransient Current Level Software Current LimitTransient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Triggered Voltage Level Transient Voltage LevelTransient Operation Continuous Transient OperationPulsed Transient Operation Toggled Transient OperationTriggering a preset level List OperationTriggered Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Protection Features Resetting Latched ProtectionInput Control Short On/OffOverpower OvertemperatureGeneral Information Overvoltage OvercurrentStatus Reporting Reading Remote Programming ErrorsReverse Voltage Saving and Recalling Settings External Control SignalsRemote Sensing Monitor OutputsPort On/Off FaultInput Measurements Power Measurements DC MeasurementsRMS Measurements Minimum and Maximum MeasurementsCleaning InspectionDamage Packaging Material Items SuppliedInstallation Installing the ModulesProcedure Location Channel Number3A. N3300A Outline Diagram Do not block the fan exhaust at the rear of the unit Bench OperationRack Mounting Input Connections Power CordManually-Tightened Connectors 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Maximum Wire Lengths to Limit Voltage Drops Stranded Copper Wire Ampere CapacityWire Size Ampacity Control Connector A1-A3A11 and A10 +S and -SSense Switch Trigger and Digital ConnectionsRS-232 Interface Computer ConnectionsGpib Interface Application Connections Local Sense ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingCheckout Programming Values Checkout ProcedureIntroduction Error Messages Selftest ErrorsSelftest Error Codes Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysDisplay Measurement Scrolling KeysMetering Keys Input Control Keys InpshorList Control Keys Transient Control KeysTrigger Control Keys Entry Keys Entry KeysUsing the Front Panel Display Set the constant current mode inputExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Querying and Clearing Output Protection and Errors Query and Clear Errors as followsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsSetting the Gpib Address Storing and Recalling Instrument StatesSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredCC Mode Tests Performance TestsImon Zero Verification Performance Test and CalibrationCV Mode Tests CC Mode Regulation TestPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test Resistance Range 2 Programming Accuracy CR Mode TestsResistance Range 1 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Parameters Calibrated Table B-2 Variables and Power Source SettingsCalibration IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 United States Latin America Canada Australia/New ZealandEurope Asia Pacific JapanManual Updates