Agilent Technologies n3303a, n3305a manual Performance Tests, Imon Zero Verification, CC Mode Tests

Page 73

B - Performance Test and Calibration

Performance Tests

IMON Zero Verification

This test verifies that IMON Zero is within calibration. IMON zero must be in calibration to perform the tests in this section. If the IMON value is not within calibration, go to the calibration section of this appendix and calibrate IMON zero, IPROG, current programming and readback and resistance programming and readback. See figure B-1 for IMON Zero test setup.

Action

1.Turn off load module and connect DVM to IMON as per figure B-1. Input terminals must be open ( no power source connected )

2.Turn on Load – press [ Recall } – scroll to *RST – press [Enter ]

3.Read IMON voltage from DVM.

4.If IMON in step 3 in not within specification then calibrate IMON zero, IPROG, Current mode and Resistance mode.

Normal Result

Voltage reading less than 0.166mV for models N3302A through N3305A Voltage reading less than 0.320mV for model N3306A

CC Mode Tests

These tests verify that the module operates in the CC mode and that IMON, current programming, and readback to the front panel display are within specification. Values read back over the GPIB are the same as those displayed on the front panel. To read back the current value via the GPIB, use MEAS:CURR:ACDC?

Current Programming and Measurement Accuracy Test

This test verifies that the current programming and measurement accuracy are within specification. If the test readings are out of tolerance the module may require calibration. If the test readings significantly disagree with the specified values or no readings can be obtained go to the Turn-On Checkout procedure to verify module operation.

Make a copy of the module test card for the model to be tested to record test values.

1.

Connect the Electronic Load, power source, DVM and current shunt as shown in figure B-2.

 

 

 

2.

Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].

3.

Turn on the power source. Set the power source voltage and current to the values listed in the following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

5V

 

20V

 

5V

8V

5V

 

 

Power Source Current Setting

33A

 

12A

 

61A

61A

130A

 

4.

Checking high current range, Low input current. Press [ CURR ] [ 1 ] then [ Enter ]. Wait 10 seconds then record the

 

actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test card under

 

high current range low current.

 

 

 

 

 

 

 

 

 

 

 

5.

Checking high current range, maximum input current. Press [ CURR ] [ select current from following table] [Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

N3305A

 

N3306A

 

 

 

 

30

10

 

60

 

 

60

 

120

 

 

 

Wait 30 seconds, then record the actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test card under high current range high current.

76

Image 73
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols Declaration of Conformity DeclarationPrinting History Acoustic Noise InformationTable of Contents TURN-ON Checkout 101 Front Panel -At a Glance Quick ReferenceRear Panel At a Glance Instrument ConfigurationQuick Reference Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance Voltage Quick Reference Front Panel MenusComplete description of all programming commands Scpi Programming Commands At a GlanceQuick Reference Scpi Commands Document Orientation Task Where to find informationGetting Started Map Options and Accessories Safety ConsiderationsDescription Front Panel Controls Features and CapabilitiesRemote Programming Operating Modes Constant Current CC ModeImmediate Current Level General Information Triggered Current Level Constant Resistance CR ModeTransient Current Level Software Current LimitTriggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Transient Voltage Level Triggered Voltage LevelContinuous Transient Operation Transient OperationPulsed Transient Operation Toggled Transient OperationList Operation Triggering a preset levelTriggered Operation Triggering a transient pulseRisetime Transition Limitation Slew Rate and Minimum Transition TimeTransition Time Slew Rate Examples Resetting Latched Protection Protection FeaturesInput Control Short On/OffOvertemperature OverpowerGeneral Information Overvoltage OvercurrentReverse Voltage Reading Remote Programming ErrorsStatus Reporting External Control Signals Saving and Recalling SettingsRemote Sensing Monitor OutputsInput Measurements FaultPort On/Off DC Measurements Power MeasurementsRMS Measurements Minimum and Maximum MeasurementsInspection CleaningDamage Packaging Material Items SuppliedProcedure Installing the ModulesInstallation Channel Number Location3A. N3300A Outline Diagram Rack Mounting Bench OperationDo not block the fan exhaust at the rear of the unit Manually-Tightened Connectors Power CordInput Connections 6a. Manual Connector 8mm Screw Terminal Connector option UJ1Wire Considerations Wire Size Ampacity Stranded Copper Wire Ampere CapacityMaximum Wire Lengths to Limit Voltage Drops A1-A3 Control ConnectorA11 and A10 +S and -STrigger and Digital Connections Sense SwitchGpib Interface Computer ConnectionsRS-232 Interface Local Sense Connections Application ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing 13. Zero Volt Loading Low Voltage OperationIntroduction Checkout ProcedureCheckout Programming Values Selftest Errors Error MessagesSelftest Error Codes Case of TroubleFront Panel, Overall View Front Panel DescriptionFront Panel Operation System Keys Display Command FunctionImmediate Action Keys Function keysMetering Keys Scrolling KeysDisplay Measurement Inpshor Input Control KeysTrigger Control Keys Transient Control KeysList Control Keys Entry Keys Entry KeysSet the constant current mode input Using the Front Panel DisplayExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant voltage mode input Set the constant resistance mode inputProgramming Transient Operation Set transient operation for Continuous modeSet transient operation for Toggled mode Set transient operation for Pulse modeTmode Togg Programming ListsList Sequence Query and Clear Errors as follows Querying and Clearing Output Protection and ErrorsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsStoring and Recalling Instrument States Setting the Gpib AddressSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Table A-2. Supplemental Characteristics Specifications aN3302A N3303A N3304A N3305A N3306A SpecificationsEquipment Required Performance Test and Calibration ProceduresPerformance Tests CC Mode TestsImon Zero Verification Performance Test and CalibrationCC Mode Regulation Test CV Mode TestsPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test Resistance Range 1 Programming Accuracy CR Mode TestsResistance Range 2 Programming Accuracy Resistance Range 4 Programming Accuracy Resistance Range 3 Programming AccuracyFrequency and Duty Cycle Accuracy Transient Generator Mode TestsOutput 705TRAN on Output 705TRIGIMM Performance Test and Calibration Pulse Width AccuracyAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Calibration Table B-2 Variables and Power Source SettingsParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Print TABXY9,15CALIBRATING High Voltage Range Voltage Calibration ProgramPerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 101 Index102 Index103 Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates