Agilent Technologies n3306a, n3305a, n3304a, n3301a, n3303a Agilent N3303A Verification Test Record

Page 81

B - Performance Test and Calibration

Agilent N3303A Verification Test Record

Test Description

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

10 Ampere Range Programming and Readback

0.9915 A

_______A

1.041 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 5.5mA

_______A

Aout + 5.5mA

High Current ( 10 A )

 

9.9825 A

_______A

10.0175 A

Front Panel Display

 

Aout – 10mA

_______A

Aout + 10mA

1 Ampere Range Programming and Readback

0.0959 A

_______A

0.1041 A

Low Current ( 0.1 A )

 

Front Panel Display

 

Aout – 2.55 mA

_______A

Aout + 2.55mA

High Current ( 1 A )

 

0.995 A

_______A

1.005 A

Front panel Display

 

Aout – 3mA

_______A

Aout + 3mA

 

 

 

 

 

 

Current Regulation test

 

 

1.0 Amp Regulation at input voltage delta 4V to 240V

 

______mA

Less then 8mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

240 Volt Range Programming and Readback

2.957 V

_______V

3.043 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 21mV

_______V

Vout + 21mV

High Voltage ( 240 V )

 

239.720 V

_______V

240.280 V

Front Panel Display

 

Vout – 140mV

_______V

Vout + 140mV

24 Volt Range Programming and Readback

2.987 V

_______V

3.013 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 11mV

_______V

Vout + 11mV

High Voltage ( 24 V )

 

23.966 V

_______V

24.034 V

Front Panel Display

 

Vout – 22mV

_______V

Vout + 22mV

 

 

 

 

 

 

Voltage Mode

Regulation Test

 

 

25V regulation with input current delta 0.4A to 10A

 

______mV

Less then 10mV

 

 

 

 

 

 

Resistance

Mode Tests

 

 

Resistance Range 1

 

 

______ Ω

49.54 Ω

High Resistance Point

48 Ω

46.46 Ω

Low Resistance Point

1.2 Ω

1.088 Ω

______ Ω

1.312 Ω

Resistance Range 2

 

41.30 Ω

______Ω

46.70 Ω

Low Resistance Point

44 Ω

High Resistance Point

480 Ω

383.50 Ω

______Ω

576.5 Ω

Resistance Range 3

 

325 Ω

______Ω

598.4 Ω

Low Resistance Point

440 Ω

High Resistance Point

4800 Ω

2640 Ω

______Ω

38,400 Ω

Resistance Range 4

 

1496 Ω

______Ω

39,600 Ω

Low Resistance Point

4400 Ω

High Resistance Point

12,000 Ω

4080 Ω

______Ω

108,000 Ω

 

 

 

 

 

 

Transient Generator Tests

 

 

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

 

84

Image 81
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols Declaration of Conformity DeclarationPrinting History Acoustic Noise InformationTable of Contents TURN-ON Checkout 101 Front Panel -At a Glance Quick ReferenceInstrument Configuration Rear Panel At a GlanceQuick Reference Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance Voltage Quick Reference Front Panel MenusComplete description of all programming commands Scpi Programming Commands At a GlanceQuick Reference Scpi Commands Task Where to find information Document OrientationGetting Started Map Safety Considerations Options and AccessoriesDescription Features and Capabilities Front Panel ControlsRemote Programming Constant Current CC Mode Operating ModesImmediate Current Level General Information Triggered Current Level Constant Resistance CR ModeTransient Current Level Software Current LimitConstant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Transient Voltage Level Triggered Voltage LevelContinuous Transient Operation Transient OperationPulsed Transient Operation Toggled Transient OperationList Operation Triggering a preset levelTriggered Operation Triggering a transient pulseRisetime Transition Limitation Slew Rate and Minimum Transition TimeTransition Time Slew Rate Examples Resetting Latched Protection Protection FeaturesInput Control Short On/OffOvertemperature OverpowerGeneral Information Overvoltage OvercurrentReading Remote Programming Errors Reverse VoltageStatus Reporting External Control Signals Saving and Recalling SettingsRemote Sensing Monitor OutputsFault Input MeasurementsPort On/Off DC Measurements Power MeasurementsRMS Measurements Minimum and Maximum MeasurementsInspection CleaningDamage Packaging Material Items SuppliedInstalling the Modules ProcedureInstallation Channel Number Location3A. N3300A Outline Diagram Bench Operation Rack MountingDo not block the fan exhaust at the rear of the unit Power Cord Manually-Tightened ConnectorsInput Connections 6a. Manual Connector 8mm Screw Terminal Connector option UJ1Wire Considerations Stranded Copper Wire Ampere Capacity Wire Size AmpacityMaximum Wire Lengths to Limit Voltage Drops A1-A3 Control ConnectorA11 and A10 +S and -STrigger and Digital Connections Sense SwitchComputer Connections Gpib InterfaceRS-232 Interface Local Sense Connections Application ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing 13. Zero Volt Loading Low Voltage OperationCheckout Procedure IntroductionCheckout Programming Values Selftest Errors Error MessagesSelftest Error Codes Case of TroubleFront Panel, Overall View Front Panel DescriptionFront Panel Operation System Keys Display Command FunctionImmediate Action Keys Function keysScrolling Keys Metering KeysDisplay Measurement Inpshor Input Control KeysTransient Control Keys Trigger Control KeysList Control Keys Entry Keys Entry KeysSet the constant current mode input Using the Front Panel DisplayExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant voltage mode input Set the constant resistance mode inputProgramming Transient Operation Set transient operation for Continuous modeSet transient operation for Toggled mode Set transient operation for Pulse modeTmode Togg Programming ListsList Sequence Query and Clear Errors as follows Querying and Clearing Output Protection and ErrorsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsStoring and Recalling Instrument States Setting the Gpib AddressSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Table A-2. Supplemental Characteristics Specifications aN3302A N3303A N3304A N3305A N3306A SpecificationsEquipment Required Performance Test and Calibration ProceduresPerformance Tests CC Mode TestsImon Zero Verification Performance Test and CalibrationCC Mode Regulation Test CV Mode TestsPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test CR Mode Tests Resistance Range 1 Programming AccuracyResistance Range 2 Programming Accuracy Resistance Range 4 Programming Accuracy Resistance Range 3 Programming AccuracyFrequency and Duty Cycle Accuracy Transient Generator Mode TestsOutput 705TRAN on Output 705TRIGIMM Performance Test and Calibration Pulse Width AccuracyAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Table B-2 Variables and Power Source Settings CalibrationParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Print TABXY9,15CALIBRATING High Voltage Range Voltage Calibration ProgramPerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 101 Index102 Index103 Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates