Agilent Technologies n3305a, n3304a Specifications a, Table A-2. Supplemental Characteristics

Page 70

Specifications - A

Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical performance determined either by design or type testing.

Table A-2. Supplemental Characteristics

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

Programming Resolution

 

 

 

 

 

 

 

 

 

 

 

Constant Current Mode

0.05mA/0.5mA

 

0.02mA/0.2mA

0.1 mA/1mA

0.1mA/1mA

0.2 mA / 2 mA

Constant Voltage Mode

0.1mV/1mV

 

0.4 mV/4mV

 

0.1mV/1mV

0.25mV/2.5mV

0.1mV / 1mV

Constant Resistance Mode

0.07/0.7/7/70mΩ

 

0.82/8.2/82mΩ

0.035/0.35/3.5/

0.085/0.85/8.5/

0.0175/0.175/

 

 

 

 

 

 

 

35mΩ

85mΩ

1.75/17.5mΩ

Readback Resolution

 

 

 

 

 

 

 

 

 

 

 

Current

 

0.05mA/0.5mA

 

0.02mA/0.2mA

0.1 mA / 1mA

0.1mA / 1 mA

0.2 mA / 2 mA

Voltage

 

0.1mV / 1mV

 

0.4 mV / 4mV

 

0.1 mV / 1mV

0.25mV/2.5mV

0.1mV / 1mV

Programmable Slew Rate1

 

 

 

 

 

 

 

 

 

 

 

Current Ranges

Slow band

500A/s - 25kA/s

 

167A/s - 8330A/s

1kA/s - 50kA/s

1kA/s - 50kA/s

2kA/s - 100kA/s

Fast band ≥3V

50kA/s - 2.5MA/s

 

16.7kA/s - 833kA/s

100kA/s - 5MA/s

100kA/s - 5MA/s

200kA/s - 10MA/s

Fast band <3V

50kA/s - 250kA/s

 

16.7kA/s - 83.3kA/s

100kA/s - 500kA/s

100kA/s - 500kA/s

200kA/s - 1MA/s

 

 

 

 

 

 

 

 

 

 

 

Voltage Ranges

Slow band

1kV/s - 50kV/s

 

4kV/s - 200kV/s

 

1kV/s - 50kV/s

2.5kV/s - 125kV/s

1kV/s - 50kV/s

Fast band ≥3V

100kV/s - 500kV/s

 

400kV/s - 2MV/s

100kV/s - 500kV/s

250kV/s -1.25MV/s

100kV/s - 500kV/s

Fast band <3V

100kV/s - 50kV/s

 

400kV/s - 200kV/

100kV/s - 50kV/s

250kV/s -125kV/s

100kV/s - 50kV/s

 

 

 

 

 

 

 

 

 

 

 

Resistance Range 1

Slow band

44Ω/s - 1125Ω/s

 

540Ω/s - 13.5kΩ/s

22Ω/s - 560Ω/s

55Ω/s - 1400Ω/s

11Ω/s - 280Ω/s

 

2250Ω/s - 34kΩ/s

 

27kΩ/s - 408kΩ/s

1120Ω/s - 17kΩ/s

2800Ω/s - 42.5kΩ/s

560Ω/s - 8.5kΩ/s

Fast band 3V

2250Ω/s - 3.4kΩ/s

 

27kΩ/s - 40.8kΩ/s

1120Ω/s - 1.7kΩ/s

2800Ω/s - 4.25kΩ/s

560Ω/s - 850Ω/s

Fast band <3V

 

Resistance Range 2

Slow band

440Ω/s - 11.25kΩ/s

 

5.4kΩ/s - 135kΩ/s

220Ω/s - 5600Ω/s

550Ω/s - 14kΩ/s

110Ω/s - 2800Ω/s

 

22.5kΩ/s - 340kΩ/s

 

270kΩ/s -4.08MΩ/s

11.2kΩ/s - 170kΩ/s

28kΩ/s - 425kΩ/s

5600Ω/s - 85kΩ/s

Fast band 3V

22.5kΩ/s - 34kΩ/s

 

270kΩ/s -408kΩ/s

11.2kΩ/s - 17kΩ/s

28kΩ/s - 42.5kΩ/s

5600Ω/s - 8.5kΩ/s

Fast band <3V

 

Resistance Range 3

Slow band

4.4kΩ/s - 112.5kΩ/s

 

54kΩ/s - 1.35MΩ/s

2.2kΩ/s - 56kΩ/s

5.5kΩ/s - 140kΩ/s

1.1kΩ/s - 28kΩ/s

 

225kΩ/s - 3.4MΩ/s

 

2.7MΩ/s-40.8MΩ/s

112kΩ/s - 1.7M Ω/s

280kΩ/s -4.25MΩ/s

56kΩ/s - 850kΩ/s

Fast band 3V

225kΩ/s - 340kΩ/s

 

2.7MΩ/s-4.08MΩ/s

112kΩ/s - 170k Ω/s

280kΩ/s -425kΩ/s

56kΩ/s - 85kΩ/s

Fast band <3V

 

Resistance Range 4

Slow band

44kΩ/s - 1.125MΩ/s

 

540kΩ/s -13.5MΩ/s

22kΩ/s - 560kΩ/s

55kΩ/s - 1.4MΩ/s

11kΩ/s - 280kΩ/s

 

2.25MΩ/s - 34MΩ/s

 

27MΩ/s - 408MΩ/s

1.12MΩ/s -17MΩ/s

2.8MΩ/s-42.5MΩ/s

560kΩ/s - 8.5MΩ/s

Fast band 3V

2.25MΩ/s - 3.4MΩ/s

 

27MΩ/s -40.8MΩ/s

1.12MΩ/s-1.7MΩ/s

2.8MΩ/s-4.25MΩ/s

560kΩ/s - 850kΩ/s

Fast band <3V

 

Slew Rate Accuracy

 

 

 

 

within

35% of programmed

value

 

 

 

Programmable Short

66m

max.

 

200m max.

 

33m

max.

33m

max.

17m

max.

Ω

 

Ω

 

Ω

Ω

Ω

 

 

40mΩ typical

 

100mΩ typical

20mΩ typical

25mΩ typical

12mΩ typical

Programmable Open

 

20k

 

80k

 

20k

20k

20k

 

Ω

 

≥ Ω

 

Ω

Ω

Ω

DC Isolation Voltage

 

 

 

 

±300 Vdc between +

or - input binding post

and chassis ground

 

 

Command Processing Time

 

 

 

 

 

 

 

 

 

 

 

Using discrete commands

 

 

 

 

 

 

3 ms

 

 

 

 

Using List commands

 

 

 

 

 

 

 

1 ms

 

 

 

 

List Dwell Characteristics

 

 

 

 

 

 

 

 

 

 

 

Range

 

 

 

 

 

 

 

0 - 10 s

 

 

 

 

Resolution

 

 

 

 

 

 

 

1 ms

 

 

 

 

Accuracy

 

 

 

 

 

 

 

5 ms

 

 

 

 

Measurement Time

 

 

 

 

 

 

 

 

 

 

 

 

1000 samples (default)

 

 

 

20 ms (with specified measurement accuracy)

 

 

200 samples

 

 

 

 

10 ms (with < 6% additional fixed offset)

 

 

100 samples

 

 

 

 

9 ms

(with < 10% additional fixed offset)

 

 

20 points

 

 

 

 

7 ms

(with < 30% additional fixed offset)

 

 

< 20 points

 

 

 

 

7 ms

(with > 30% additional fixed offset)

 

 

Measurement Bandwidth

 

 

 

 

 

10kHz (rms)

 

 

 

 

GPIB Capabilities

 

 

 

 

SH1, AH1, T6, L4, SR1, RL1, DT1, CD1

 

 

1Slew rate bands are not programmable. When you program a slew rate value outside the indicated bands, the electronic load will automatically adjust the slew rate to fit within the band that is closest to the programmed value. The slew rate accuracy specification is only applicable to slew rates within the indicated bands. Additionally, when the transition from one setting to another is small, the small signal bandwidth of the load limits the minimum transition time for all programmable slew rates (see chapter 2 for more information).

Below 3 volts, the maximum bandwidth of the electronic load is reduced by a factor of ten to one. For example, in the current range for Model N3302A, the maximum slew rate is specified as 2.5MA/s, below 3 volts the maximum slew rate would be 250kA/s. Any slew rate programmed between 2.5MA/s and 250kA/s would produce a slew rate of 250kA/s. Slew rates programmed slower than 250kA/s would still correctly reflect their programmed value. If you are using transient mode to generate a high frequency pulse train, a reduced slew rate might cause the load to never reach the upper programmed value before beginning the transition to the lower programmed value. So even though the transient mode is still operational at lower voltages, a fast pulse train with large transitions may not be achievable.

73

Image 70
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceRear Panel At a Glance Instrument ConfigurationQuick Reference Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Document Orientation Task Where to find informationGetting Started Map Options and Accessories Safety ConsiderationsDescription Front Panel Controls Features and CapabilitiesRemote Programming Operating Modes Constant Current CC ModeImmediate Current Level Transient Current Level Constant Resistance CR ModeGeneral Information Triggered Current Level Software Current LimitTriggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Triggered Voltage Level Transient Voltage LevelPulsed Transient Operation Transient OperationContinuous Transient Operation Toggled Transient OperationTriggered Operation Triggering a preset levelList Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Input Control Protection FeaturesResetting Latched Protection Short On/OffGeneral Information Overvoltage OverpowerOvertemperature OvercurrentReverse Voltage Reading Remote Programming ErrorsStatus Reporting Remote Sensing Saving and Recalling SettingsExternal Control Signals Monitor OutputsInput Measurements FaultPort On/Off RMS Measurements Power MeasurementsDC Measurements Minimum and Maximum MeasurementsDamage Packaging Material CleaningInspection Items SuppliedProcedure Installing the ModulesInstallation Location Channel Number3A. N3300A Outline Diagram Rack Mounting Bench OperationDo not block the fan exhaust at the rear of the unit Manually-Tightened Connectors Power CordInput Connections 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Wire Size Ampacity Stranded Copper Wire Ampere CapacityMaximum Wire Lengths to Limit Voltage Drops A11 and A10 Control ConnectorA1-A3 +S and -SSense Switch Trigger and Digital ConnectionsGpib Interface Computer ConnectionsRS-232 Interface Remote Sense Connections Application ConnectionsLocal Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingIntroduction Checkout ProcedureCheckout Programming Values Selftest Error Codes Error MessagesSelftest Errors Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysMetering Keys Scrolling KeysDisplay Measurement Input Control Keys InpshorTrigger Control Keys Transient Control KeysList Control Keys Entry Keys Entry KeysExamples of Front Panel Programming Using the Front Panel DisplaySet the constant current mode input Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Making Basic Front Panel Measurements Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows Use the Meter menu for making front panel measurementsSet the Gpib address as follows Setting the Gpib AddressStoring and Recalling Instrument States Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredImon Zero Verification CC Mode TestsPerformance Tests Performance Test and CalibrationPerformance Test and Calibration B CV Mode TestsCC Mode Regulation Test Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test Resistance Range 1 Programming Accuracy CR Mode TestsResistance Range 2 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Calibration Table B-2 Variables and Power Source SettingsParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 Europe Asia Pacific United States Latin AmericaCanada Australia/New Zealand JapanManual Updates