Performance Test and Calibration - B

Agilent N3306A Verification Test Record

Test Description

 

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

120 Ampere Range Programming and Readback

0.9615 A

_______A

1.0385 A

Low Current ( 1A )

 

Front Panel Display

 

Aout – 20.5mA

_______A

Aout + 20.5mA

High Current ( 120 A )

 

119.8425 A

_______A

120.1575 A

Front Panel Display

 

Aout – 80mA

_______A

Aout + 80mA

12 Ampere Range Programming and Readback

0.9840A

_______A

1.0160 A

Low Current ( 1A )

 

Front Panel Display

 

Aout – 10.5mA

_______A

Aout + 10.5mA

High Current ( 12 A )

 

11.973 A

_______A

12.027 A

Front panel Display

 

Aout – 16mA

_______A

Aout + 16mA

 

 

 

 

 

 

Current Regulation Test

 

 

10 Amp Regulation at input voltage delta 4V to 60V

 

______mA

Less then 10mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

60 Volt Range Programming and Readback

 

2.989 V

_______V

3.011 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 21mV

_______V

Vout + 21mV

High Voltage ( 60 V )

 

59.932 V

_______V

60.068 V

Front Panel Display

 

Vout – 38mV

_______V

Vout + 38mV

6 Volt Range Programming and Readback

 

2.9905 V

_______V

3.0095 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 4.5m

_______V

Vout + 4.5m

High Voltage ( 6 V )

 

5.991 V

_______V

6.009 V

Front Panel Display

 

Vout – 6mV

_______V

Vout + 6mV

 

 

 

 

 

 

Voltage Regulation Test

 

 

5V regulation with input current delta 1A to 120A

 

______mV

Less then 20mV

 

 

 

 

 

Resistance Mode Tests

Resistance Range 1

 

High Resistance Point

1 Ω

Low Resistance Point

0.100 Ω

Resistance Range 2

 

Low Resistance Point

0.9 Ω

High Resistance Point

10 Ω

Resistance Range 3

 

Low Resistance Point

9 Ω

High Resistance Point

100 Ω

Resistance Range 4

 

Low Resistance Point

90 Ω

High Resistance Point

1000 Ω

0.990 Ω

0.0936 Ω

0.8665 Ω

9.680 Ω

8.400 Ω

79.940 Ω

72

500 Ω

______ Ω

______ Ω

______Ω

______Ω

______Ω

______Ω

______Ω

______ Ω

1.010 Ω

0.1064 Ω

0.9335 Ω

10.32

9.60

120.06 Ω

149.4

21,000 Ω

Transient Generator Tests

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

87

Image 84
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceInstrument Configuration Rear Panel At a GlanceQuick Reference Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Task Where to find information Document OrientationGetting Started Map Safety Considerations Options and AccessoriesDescription Features and Capabilities Front Panel ControlsRemote Programming Constant Current CC Mode Operating ModesImmediate Current Level Constant Resistance CR Mode General Information Triggered Current LevelTransient Current Level Software Current LimitConstant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Triggered Voltage Level Transient Voltage LevelTransient Operation Continuous Transient OperationPulsed Transient Operation Toggled Transient OperationTriggering a preset level List OperationTriggered Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Protection Features Resetting Latched ProtectionInput Control Short On/OffOverpower OvertemperatureGeneral Information Overvoltage OvercurrentReading Remote Programming Errors Reverse VoltageStatus Reporting Saving and Recalling Settings External Control SignalsRemote Sensing Monitor OutputsFault Input MeasurementsPort On/Off Power Measurements DC MeasurementsRMS Measurements Minimum and Maximum MeasurementsCleaning InspectionDamage Packaging Material Items SuppliedInstalling the Modules ProcedureInstallation Location Channel Number3A. N3300A Outline Diagram Bench Operation Rack MountingDo not block the fan exhaust at the rear of the unit Power Cord Manually-Tightened ConnectorsInput Connections 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Stranded Copper Wire Ampere Capacity Wire Size AmpacityMaximum Wire Lengths to Limit Voltage Drops Control Connector A1-A3A11 and A10 +S and -SSense Switch Trigger and Digital ConnectionsComputer Connections Gpib InterfaceRS-232 Interface Application Connections Local Sense ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingCheckout Procedure IntroductionCheckout Programming Values Error Messages Selftest ErrorsSelftest Error Codes Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysScrolling Keys Metering KeysDisplay Measurement Input Control Keys InpshorTransient Control Keys Trigger Control KeysList Control Keys Entry Keys Entry KeysUsing the Front Panel Display Set the constant current mode inputExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Querying and Clearing Output Protection and Errors Query and Clear Errors as followsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsSetting the Gpib Address Storing and Recalling Instrument StatesSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredCC Mode Tests Performance TestsImon Zero Verification Performance Test and CalibrationCV Mode Tests CC Mode Regulation TestPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test CR Mode Tests Resistance Range 1 Programming AccuracyResistance Range 2 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Table B-2 Variables and Power Source Settings CalibrationParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 United States Latin America Canada Australia/New ZealandEurope Asia Pacific JapanManual Updates