Agilent Technologies n3306a, n3305a, n3304a manual CV Mode Tests, Performance Test and Calibration B

Page 74

Performance Test and Calibration - B

6.Press [ CURR ] [ 1 ] ( except N3303A press [ CURR ] [ 0.1 ] ). Press [ CURR ]. Scroll until display reads C:RANG, press [ 1 ] [ Enter ]. Press shift key, [ Sense ] scroll till display reads S:C:RNG press [ 1 ] [ Enter ].

7.Checking low current range, low input current Wait 10 seconds then record the actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test card under low current range low current.

8.Checking low current range, maximum input current. Press [ CURR ] [ select current from following table ] [ Enter ].

N3302A

N3303A

N3304A

N3305A

N3306A

3

1

6

6

12

Wait 30 seconds then record the actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test card under high current range high current.

CC Mode Regulation Test

This test verifies the Input Current remains within specification when the input voltage is changed from a low voltage to rated voltage.

1.

Connect the Electronic Load, power source, DVM and current shunt as shown in figure B-2.

 

 

2.

Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ]

3.

Press [ CURR ] [ select current from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

N3305A

 

N3306A

 

 

 

2.5

1

 

5

 

 

3.3

 

10

 

 

4.

Turn on the power source. Set the power source voltage and current to the values listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

4V

4V

 

4V

4V

4V

 

Power Source Current Setting

5A

5A

 

10A

10A

20A

5.

Wait 10 seconds, then record input current reading

( DVM reading / current shunt ).

 

 

6.

Reset the power source voltage level to value listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

60V

240V

 

60V

150V

60V

7.

Wait 10 seconds, then record input current reading ( DVM reading / shunt resistance ).

 

 

8.

Subtract reading in step ‘7’ from step ‘5’. Absolute value of difference should be less then specification.

CV Mode Tests

These tests verify that the module operates in the CV mode and that voltage programming and readback to the front panel display are within specification. Values read back over the GPIB are the same as those displayed on the front panel. To read back the voltage value via the GPIB, use the command MEAS:VOLT:ACDC?

Voltage Programming and Measurement Accuracy Test

This test verifies that voltage programming and voltage measurement accuracy are within specification. Make a copy of the module test card for the model being tested to record test values.

1.Connect the Electronic Load, Power source and DVM as shown in figure B-3. Be sure + sense and – sense are connected to the +/- input terminals and the remote/local button is in the remote position.

2.Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].

3.Press [ Func ]. Scroll until display reads FUNC VOLT then press [ Enter ].

77

Image 74
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceQuick Reference Instrument ConfigurationRear Panel At a Glance Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Getting Started Map Task Where to find informationDocument Orientation Description Safety ConsiderationsOptions and Accessories Remote Programming Features and CapabilitiesFront Panel Controls Immediate Current Level Constant Current CC ModeOperating Modes Transient Current Level Constant Resistance CR ModeGeneral Information Triggered Current Level Software Current LimitTransient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Triggered Voltage Level Transient Voltage LevelPulsed Transient Operation Transient OperationContinuous Transient Operation Toggled Transient OperationTriggered Operation Triggering a preset levelList Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Input Control Protection FeaturesResetting Latched Protection Short On/OffGeneral Information Overvoltage OverpowerOvertemperature OvercurrentStatus Reporting Reading Remote Programming ErrorsReverse Voltage Remote Sensing Saving and Recalling SettingsExternal Control Signals Monitor OutputsPort On/Off FaultInput Measurements RMS Measurements Power MeasurementsDC Measurements Minimum and Maximum MeasurementsDamage Packaging Material CleaningInspection Items SuppliedInstallation Installing the ModulesProcedure Location Channel Number3A. N3300A Outline Diagram Do not block the fan exhaust at the rear of the unit Bench OperationRack Mounting Input Connections Power CordManually-Tightened Connectors 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Maximum Wire Lengths to Limit Voltage Drops Stranded Copper Wire Ampere CapacityWire Size Ampacity A11 and A10 Control ConnectorA1-A3 +S and -SSense Switch Trigger and Digital ConnectionsRS-232 Interface Computer ConnectionsGpib Interface Remote Sense Connections Application ConnectionsLocal Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingCheckout Programming Values Checkout ProcedureIntroduction Selftest Error Codes Error MessagesSelftest Errors Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysDisplay Measurement Scrolling KeysMetering Keys Input Control Keys InpshorList Control Keys Transient Control KeysTrigger Control Keys Entry Keys Entry KeysExamples of Front Panel Programming Using the Front Panel DisplaySet the constant current mode input Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Making Basic Front Panel Measurements Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows Use the Meter menu for making front panel measurementsSet the Gpib address as follows Setting the Gpib AddressStoring and Recalling Instrument States Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredImon Zero Verification CC Mode TestsPerformance Tests Performance Test and CalibrationPerformance Test and Calibration B CV Mode TestsCC Mode Regulation Test Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test Resistance Range 2 Programming Accuracy CR Mode TestsResistance Range 1 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Parameters Calibrated Table B-2 Variables and Power Source SettingsCalibration IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 Europe Asia Pacific United States Latin AmericaCanada Australia/New Zealand JapanManual Updates