B - Performance Test and Calibration
540 !
550INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1
560!
570OUTPUT @Ld;"CAL:DATA";Vp1
580!
590OUTPUT @Ld;"CAL:LEV P2"
600WAIT 1
610!
620INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2
630!
640OUTPUT @Ld;"CAL:DATA";Vp2
650!
660OUTPUT @Ld;"CAL:SAVE"
670OUTPUT @Ld;"CAL:STAT OFF"
680!
690CLEAR SCREEN
700PRINT "Voltage calibration adjustments complete"
710PRINT "Verify Voltage Adjustments"
720!
730END