Performance Test and Calibration - B

VOLTAGE Calibration Program

10! This program calibrates the voltage mode for load modules

20! N3302A, N3303A, N3304A, N3305A and N3306A.

30!

40! last update 9/6/00

50!

60! Edit the interface address on line 100 if different then 705

70!

80ASSIGN @Ld TO 705

90!

100OUTPUT @Ld;"*RST"

110OUTPUT @Ld;"CAL:STAT ON"

120OUTPUT @Ld;"FUNC VOLT"

130OUTPUT @Ld;"VOLT:RANG MAX"

140OUTPUT @Ld;"SENS:VOLT:RANG MAX"

150OUTPUT @Ld;"INP:STAT ON"

160!

170

!

*******

Calibrating high volt range

*******

180

!

 

 

 

190PRINT TABXY(10,8),"See figure B-3 for voltage calibration"

200PRINT TABXY(10,10),"Connect power leads to input terminals"

210PRINT TABXY(10,12),"Connect + sense to + input terminal"

220PRINT TABXY(10,14),"Connect - sense to - input terminal"

230PRINT TABXY(10,16),"Connect DVM to +/- sense terminals"

240PRINT TABXY(10,18),"See Table B-2 Voltage Mode Voltage / Current Settings"

250PRINT TABXY(10,20),"Set power source to voltage and current for model to be calibrated"

260PRINT TABXY(10,24),"BE SURE REMOTE BUTTON ON REAR PANEL IS IN REMOTE POSITION"

270DISP "Press CONT when ready to calibrate voltage "

280PAUSE

290CLEAR SCREEN

300!

310PRINT TABXY(9,15);"CALIBRATING HIGH VOLTAGE RANGE"

320OUTPUT @Ld;"CAL:LEV P1"

330WAIT 1

340INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1

350!

360OUTPUT @Ld;"CAL:DATA";Vp1

370!

380OUTPUT @Ld;"CAL:LEV P2"

390WAIT 1

400INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2

410!

420OUTPUT @Ld;"CAL:DATA";Vp2

430!

440

!

*******

Calibrating low voltage range *******

450

!

 

 

460

OUTPUT @Ld;"VOLT:RANG MIN"

470OUTPUT @Ld;"SENS:VOLT:RANG MIN"

480!

490CLEAR SCREEN

500PRINT TABXY(9,15);"CALIBRATING LOW VOLTAGE RANGE"

510!

520OUTPUT @Ld;"CAL:LEV P1"

530WAIT 1

95

Image 92
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceQuick Reference Instrument ConfigurationRear Panel At a Glance Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Getting Started Map Task Where to find informationDocument Orientation Description Safety ConsiderationsOptions and Accessories Remote Programming Features and CapabilitiesFront Panel Controls Immediate Current Level Constant Current CC ModeOperating Modes Constant Resistance CR Mode General Information Triggered Current LevelTransient Current Level Software Current LimitTransient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Triggered Voltage Level Transient Voltage LevelTransient Operation Continuous Transient OperationPulsed Transient Operation Toggled Transient OperationTriggering a preset level List OperationTriggered Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Protection Features Resetting Latched ProtectionInput Control Short On/OffOverpower OvertemperatureGeneral Information Overvoltage OvercurrentStatus Reporting Reading Remote Programming ErrorsReverse Voltage Saving and Recalling Settings External Control SignalsRemote Sensing Monitor OutputsPort On/Off FaultInput Measurements Power Measurements DC MeasurementsRMS Measurements Minimum and Maximum MeasurementsCleaning InspectionDamage Packaging Material Items SuppliedInstallation Installing the ModulesProcedure Location Channel Number3A. N3300A Outline Diagram Do not block the fan exhaust at the rear of the unit Bench OperationRack Mounting Input Connections Power CordManually-Tightened Connectors 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Maximum Wire Lengths to Limit Voltage Drops Stranded Copper Wire Ampere CapacityWire Size Ampacity Control Connector A1-A3A11 and A10 +S and -SSense Switch Trigger and Digital ConnectionsRS-232 Interface Computer ConnectionsGpib Interface Application Connections Local Sense ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingCheckout Programming Values Checkout ProcedureIntroduction Error Messages Selftest ErrorsSelftest Error Codes Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysDisplay Measurement Scrolling KeysMetering Keys Input Control Keys InpshorList Control Keys Transient Control KeysTrigger Control Keys Entry Keys Entry KeysUsing the Front Panel Display Set the constant current mode inputExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Querying and Clearing Output Protection and Errors Query and Clear Errors as followsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsSetting the Gpib Address Storing and Recalling Instrument StatesSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredCC Mode Tests Performance TestsImon Zero Verification Performance Test and CalibrationCV Mode Tests CC Mode Regulation TestPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test Resistance Range 2 Programming Accuracy CR Mode TestsResistance Range 1 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Parameters Calibrated Table B-2 Variables and Power Source SettingsCalibration IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 United States Latin America Canada Australia/New ZealandEurope Asia Pacific JapanManual Updates