Agilent Technologies n3300a, n3305a, n3304a, n3301a, n3303a, n3306a, n3302a manual Wait

Page 90

 

 

 

Performance Test and Calibration - B

1110

!

 

 

1120

OUTPUT @Ld;"CAL:IPR:LEV P4"

 

1130

!

 

 

1140

PRINT TABXY(20,20),"Calibrating IMON & IPROG P4 - 30 SECOND WAIT"

1150

WAIT 30

 

 

1160

CLEAR SCREEN

 

 

1170

!

 

 

1180

INPUT "Measure voltage across current shunt and enter in volts",Ip4

1190

!

 

 

1200

INPUT "Measure Imon voltage and enter in volts",Vimonp4

1210

!

 

 

1220

INPUT "Measure Iprog voltage and enter in volts",Viprogp4

1230

!

 

 

1240

Ip4=Ip4/Rshunt

 

1250

!

 

 

1260

OUTPUT @Ld;"CAL:DATA ";Ip4,Vimonp4,Viprogp4

 

1270

!

 

 

1280

! *******

Calibrating Main Current DAC

********

1290

!

 

 

1300

PRINT TABXY(10,10),"Disconnect instruments from IMON and IPROG terminals"

1310

DISP "Press CONT when ready to calibrate main current ADC"

1320

PAUSE

 

 

1330

!

 

 

1340

CLEAR SCREEN

 

 

1350

OUTPUT @Ld;"CAL:LEV P1"

 

1360

PRINT TABXY(10,20),"Calibrating high current range P1 - 30 SECOND WAIT"

1370

WAIT 30

 

 

1380

CLEAR SCREEN

 

 

1390

!

 

 

1400

INPUT "Measure voltage across current shunt and enter in volts",Ip1

1410

!

 

 

1420

Ip1=Ip1/Rshunt

 

1430

!

 

 

1440

OUTPUT @Ld;"CAL:DATA ";Ip1￿￿1450 !￿￿￿1460

OUTPUT @Ld;"CAL:LEV P2"

1470

!

 

 

1480

PRINT TABXY(10,10),"Calibrating high current range P2 - 30 SECOND WAIT"

1490

WAIT 30

 

 

1500

CLEAR SCREEN

 

 

1510

!

 

 

1520

INPUT "Measure voltage across current shunt and enter in volts",Ip2

1530

!

 

 

1540

Ip2=Ip2/Rshunt

 

1550

!

 

 

1560

OUTPUT @Ld;"CAL:DATA ";Ip2

 

1570

!

 

 

1580

OUTPUT @Ld;"INP:STAT OFF"

 

1590

OUTPUT @Ld;"CURR:RANG MIN"

 

1600

OUTPUT @Ld;"SENS:CURR:RANG MIN"

 

1610

!

 

 

1620

CLEAR SCREEN

 

 

1630

!

 

 

1640

INPUT "Enter value of low range current shunt in ohms",Rshuntl

1650

DISP "You have entered ";Rshuntl;" ohms"

 

1660

INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$

1670

IF UPC$(Y$)="Y" OR Y$="" THEN 1690

 

1680

IF UPC$(Y$)="N" OR Y$="n" THEN 1620

 

93

Image 90
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceInstrument Configuration Rear Panel At a GlanceQuick Reference Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Task Where to find information Document OrientationGetting Started Map Safety Considerations Options and AccessoriesDescription Features and Capabilities Front Panel ControlsRemote Programming Constant Current CC Mode Operating ModesImmediate Current Level Transient Current Level Constant Resistance CR ModeGeneral Information Triggered Current Level Software Current LimitConstant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Triggered Voltage Level Transient Voltage LevelPulsed Transient Operation Transient OperationContinuous Transient Operation Toggled Transient OperationTriggered Operation Triggering a preset levelList Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Input Control Protection FeaturesResetting Latched Protection Short On/OffGeneral Information Overvoltage OverpowerOvertemperature OvercurrentReading Remote Programming Errors Reverse VoltageStatus Reporting Remote Sensing Saving and Recalling SettingsExternal Control Signals Monitor OutputsFault Input MeasurementsPort On/Off RMS Measurements Power MeasurementsDC Measurements Minimum and Maximum MeasurementsDamage Packaging Material CleaningInspection Items SuppliedInstalling the Modules ProcedureInstallation Location Channel Number3A. N3300A Outline Diagram Bench Operation Rack MountingDo not block the fan exhaust at the rear of the unit Power Cord Manually-Tightened ConnectorsInput Connections 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Stranded Copper Wire Ampere Capacity Wire Size AmpacityMaximum Wire Lengths to Limit Voltage Drops A11 and A10 Control ConnectorA1-A3 +S and -SSense Switch Trigger and Digital ConnectionsComputer Connections Gpib InterfaceRS-232 Interface Remote Sense Connections Application ConnectionsLocal Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingCheckout Procedure IntroductionCheckout Programming Values Selftest Error Codes Error MessagesSelftest Errors Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysScrolling Keys Metering KeysDisplay Measurement Input Control Keys InpshorTransient Control Keys Trigger Control KeysList Control Keys Entry Keys Entry KeysExamples of Front Panel Programming Using the Front Panel DisplaySet the constant current mode input Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Making Basic Front Panel Measurements Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows Use the Meter menu for making front panel measurementsSet the Gpib address as follows Setting the Gpib AddressStoring and Recalling Instrument States Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredImon Zero Verification CC Mode TestsPerformance Tests Performance Test and CalibrationPerformance Test and Calibration B CV Mode TestsCC Mode Regulation Test Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test CR Mode Tests Resistance Range 1 Programming AccuracyResistance Range 2 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Table B-2 Variables and Power Source Settings CalibrationParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 Europe Asia Pacific United States Latin AmericaCanada Australia/New Zealand JapanManual Updates