Agilent Technologies n3302a, n3305a, n3304a, n3301a, n3303a manual Performance Test and Calibration B

Page 96

 

 

 

 

Performance Test and Calibration - B

1080

Rp2=Vt2/Ip2

 

 

 

1090

OUTPUT @Ld;"CAL:DATA";Rp2

 

 

1100

OUTPUT @Ld;"INP:STAT OFF"

 

 

1110

!

 

 

 

 

1120

!

************ CALIBRATING RANGE 3

******************

 

1130

!

 

 

 

 

1140

CLEAR SCREEN

 

 

 

1150

INPUT "Enter value of current shunt for range 3 in OHMS",Rshunt3

 

1160

PRINT TABXY(10,10);"You have entered ";Rshunt3;" ohms"

 

1170

INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$

1180

IF UPC$(Y$)="Y" OR Y$="" THEN 1200

 

 

1190

IF UPC$(Y$)="N" OR Y$="n" THEN 1140

 

 

1200

CLEAR SCREEN

 

 

 

1210

!

 

 

 

 

1220

PRINT TABXY(10,12);"Set power source to voltage and current settings from"

1230

PRINT TABXY(10,13);"Table B-1 for resistance range 3 - Point P1"

 

1240

DISP "Press CONT when ready to calibrate resistance range 3"

 

1250

PAUSE

 

 

 

 

1260

CLEAR SCREEN

 

 

 

1270

PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 3 - P1 - 7 SECOND WAIT"

1280

OUTPUT @Ld;"INP:STAT ON"

 

 

1290

OUTPUT @Ld;"RES:RANG 200" !< resistance range 3 variable from table B-1

1300

OUTPUT @Ld;"CAL:LEV P1"

 

 

1310

WAIT 7

 

 

 

 

1320

INPUT "Measure voltage across current shunt - enter value in volts",Vp1

1330

Ip1=Vp1/Rshunt3

 

 

 

1340

INPUT "Measure voltage across +/- sense terminals and enter value in

 

 

volts",Vt1

 

 

 

1350

Rp1=Vt1/Ip1

 

 

 

1360

OUTPUT @Ld;"CAL:DATA";Rp1

 

 

1370

!

 

 

 

 

1380

CLEAR SCREEN

 

 

 

1390

PRINT TABXY(10,12);"Set power source to voltage and current settings from

 

Table B-2"

 

 

 

1400

PRINT TABXY(10,13);"for resistance range 3 - Point P2"

 

1410

DISP "Press Cont to continue resistance range 3 calibration"

 

1420

PAUSE

 

 

 

 

1430

OUTPUT @Ld;"CAL:LEV P2"

 

 

1440

CLEAR SCREEN

 

 

 

1450

PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 3 - P2 - 7 SECOND

WAIT"

1460

WAIT 7

 

 

 

 

1470

INPUT "Measure voltage across current shunt - enter value in volts",Vp2

1480

Ip2=Vp2/Rshunt3

 

 

 

1490

INPUT "Measure voltage across +/- sense terminals and enter value in

 

 

volts",Vt2

 

 

 

1500

Rp2=Vt2/Ip2

 

 

 

1510

OUTPUT @Ld;"CAL:DATA";Rp2

 

 

1520

OUTPUT @Ld;"INP:STAT OFF"

 

 

1530

!

 

 

 

 

1540

!

************

CALIBRATING RANGE 4

*********************

 

1550

!

 

 

 

 

1560

CLEAR SCREEN￿F1570

INPUT "Enter value of current shunt for range 4 in

 

OHMS",Rshunt4

 

 

 

1580

PRINT TABXY(10,10);"You have entered ";Rshunt4;" ohms"

 

1590

INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$

1600

IF UPC$(Y$)="Y" OR Y$="" THEN 1620

 

 

1610

IF UPC$(Y$)="N" OR Y$="n" THEN 1560

 

 

99

Image 96
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration Declaration of ConformityAcoustic Noise Information Printing HistoryTable of Contents TURN-ON Checkout 101 Quick Reference Front Panel -At a GlanceInstrument Configuration Rear Panel At a GlanceQuick Reference Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Quick Reference Front Panel Menus VoltageScpi Programming Commands At a Glance Complete description of all programming commandsQuick Reference Scpi Commands Task Where to find information Document OrientationGetting Started Map Safety Considerations Options and AccessoriesDescription Features and Capabilities Front Panel ControlsRemote Programming Constant Current CC Mode Operating ModesImmediate Current Level Constant Resistance CR Mode General Information Triggered Current LevelTransient Current Level Software Current LimitConstant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Triggered Voltage Level Transient Voltage LevelTransient Operation Continuous Transient OperationPulsed Transient Operation Toggled Transient OperationTriggering a preset level List OperationTriggered Operation Triggering a transient pulseSlew Rate and Minimum Transition Time Risetime Transition LimitationTransition Time Slew Rate Examples Protection Features Resetting Latched ProtectionInput Control Short On/OffOverpower OvertemperatureGeneral Information Overvoltage OvercurrentReading Remote Programming Errors Reverse VoltageStatus Reporting Saving and Recalling Settings External Control SignalsRemote Sensing Monitor OutputsFault Input MeasurementsPort On/Off Power Measurements DC MeasurementsRMS Measurements Minimum and Maximum MeasurementsCleaning InspectionDamage Packaging Material Items SuppliedInstalling the Modules ProcedureInstallation Location Channel Number3A. N3300A Outline Diagram Bench Operation Rack MountingDo not block the fan exhaust at the rear of the unit Power Cord Manually-Tightened ConnectorsInput Connections 8mm Screw Terminal Connector option UJ1 6a. Manual ConnectorWire Considerations Stranded Copper Wire Ampere Capacity Wire Size AmpacityMaximum Wire Lengths to Limit Voltage Drops Control Connector A1-A3A11 and A10 +S and -SSense Switch Trigger and Digital ConnectionsComputer Connections Gpib InterfaceRS-232 Interface Application Connections Local Sense ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing Low Voltage Operation 13. Zero Volt LoadingCheckout Procedure IntroductionCheckout Programming Values Error Messages Selftest ErrorsSelftest Error Codes Case of TroubleFront Panel Description Front Panel, Overall ViewFront Panel Operation Display Command Function System KeysFunction keys Immediate Action KeysScrolling Keys Metering KeysDisplay Measurement Input Control Keys InpshorTransient Control Keys Trigger Control KeysList Control Keys Entry Keys Entry KeysUsing the Front Panel Display Set the constant current mode inputExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant resistance mode input Set the constant voltage mode inputSet transient operation for Continuous mode Programming Transient OperationSet transient operation for Pulse mode Set transient operation for Toggled modeProgramming Lists Tmode ToggList Sequence Querying and Clearing Output Protection and Errors Query and Clear Errors as followsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsSetting the Gpib Address Storing and Recalling Instrument StatesSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Specifications a Table A-2. Supplemental CharacteristicsSpecifications N3302A N3303A N3304A N3305A N3306APerformance Test and Calibration Procedures Equipment RequiredCC Mode Tests Performance TestsImon Zero Verification Performance Test and CalibrationCV Mode Tests CC Mode Regulation TestPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test CR Mode Tests Resistance Range 1 Programming AccuracyResistance Range 2 Programming Accuracy Resistance Range 3 Programming Accuracy Resistance Range 4 Programming AccuracyTransient Generator Mode Tests Frequency and Duty Cycle AccuracyPerformance Test and Calibration Pulse Width Accuracy Output 705TRAN on Output 705TRIGIMMAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Table B-2 Variables and Power Source Settings CalibrationParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Voltage Calibration Program Print TABXY9,15CALIBRATING High Voltage RangePerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 Index 101Index 102103 United States Latin America Canada Australia/New ZealandEurope Asia Pacific JapanManual Updates