Agilent Technologies n3300a, n3305a, n3304a, n3301a, n3303a, n3306a, n3302a manual Specifications

Page 69

A - Specifications

Table A-1. Specifications (continued)

N3302A

N3303A

N3304A

N3305A

N3306A

Constant Current Mode2

Low Range/High Range

Regulation

Low Range Accuracy

High Range Accuracy

Constant Voltage Mode2

Low Range/High Range

Regulation

Low Range Accuracy

High Range Accuracy

Constant Resistance Mode2

Range 1 (I >10% of current rating) Accuracy up to 25% of range Accuracy >25% to 50% of range Accuracy >50% to 100% of range

Range 2 (I >1% of current rating) Accuracy up to 25% of range Accuracy >25% to 50% of range Accuracy >50% to 100% of range

Range 3 (I >0.1% of current rating) Accuracy up to 25% of range Accuracy >25% to 50% of range Accuracy >50% to 100% of range

Range 4 (I >0.01% of current rating) Accuracy up to 25% of range Accuracy >25% to 50% of range Accuracy >50% to 80% of range Accuracy >80% to 100% of range

3A / 30A

10mA

0.1% + 5mA

0.1% + 10mA

6V / 60V

5mV

0.1% + 3 mV

0.1% + 8 mV

0.067-4 Ω

0.4% ± 24mΩ

0.4% ± 24mΩ

0.8% ± 24mΩ

3.6-40 Ω

1.5% ±80mΩ

2% ±80mΩ

3% ±80mΩ

36-400 Ω

8% ±240mΩ

16% ±240mΩ

32% ±240mΩ

360-2000Ω -25/+100% -25/+100% -40/+400% -40/+400%

1A / 10A

8mA

0.1% + 4mA

0.1% + 7.5mA

24V / 240V

10mV

0.1% + 10mV

0.1% + 40mV

0.2-48 Ω

1% ± 100mΩ

2% ± 100mΩ

3% ± 100mΩ

44-480 Ω

5% ± 500mΩ

10% ± 500mΩ

20% ± 500mΩ

440-4800Ω -30/+50% -40/+133% -50/+600%

4400-12000Ω -66/+800% -66/+800% -66/+800% -66/+800%

6A / 60A

10mA

0.1% + 7.5mA

0.1% + 15mA

6V / 60V

10mV

0.1% + 3 mV

0.1% + 8mV

0.033-2 Ω

0.4% ±12mΩ

0.4% ±12mΩ

0.4% ±12mΩ

1.8-20 Ω

1.5% ±40mΩ

2% ±40mΩ

3% ±40mΩ

18-200 Ω

6% ±120mΩ

10% ±120mΩ

20% ±120mΩ

180-2000Ω -20/+66% -33/+200% -45/+600% -50/+2000%

6A / 60A

10mA

0.1% + 7.5mA

0.1% + 15mA

15V / 150V

10mV

0.1% + 10mV

0.1% + 20mV

0.033-5 Ω

0.4% ± 24mΩ

0.4% ± 24mΩ

0.8% ± 24mΩ

4.5-50 Ω

1.5% ±80mΩ

2% ±80mΩ

3% ±80mΩ

45-500Ω -10/+25% -20/+50% -30/+150%

450-2500Ω -35/+200% -45/+500% -50/+1200% -50/+2000%

12A / 120A

10mA

0.1% + 15mA

0.1% + 37.5mA

6V / 60V

20mV

0.1% + 3mV

0.1% + 8mV

0.017-1 Ω

0.4% ± 6mΩ

0.4% ± 6mΩ

0.4% ± 6mΩ

0.9-10 Ω

1.5% ±20mΩ

2% ±20mΩ

3% ±20mΩ

9-100 Ω

6% ±60mΩ

10% ±60mΩ

20% ±60mΩ

90-1000Ω -20/+66% -33/+200% -45/+600% -50/+2000%

Transient Generator

Frequency Range

Frequency Accuracy

Duty Cycle Range

Duty Cycle Accuracy3 Pulse Width

Current Measurement2

Low Range / High Range

Low Range Accuracy4

High Range Accuracy4

Voltage Measurement2

Low Range / High Range

Low Range Accuracy

High Range Accuracy

Power Measurement2

Accuracy

0.25Hz -10kHz

0.5%

3 to 97% (0.25Hz - 1kHz)

6 to 94% (1kHz - 10kHz)

1%

50μs ± 1% to 4 seconds ± 1%

3A / 30A

 

1A / 10A

 

6A / 60A

 

6A / 60A

 

12A / 120A

0.05% + 3mA

 

0.05%

+ 2.5mA

 

0.05%

+ 5mA

 

0.05%

+ 5mA

 

0.05%

+ 10mA

0.05% + 6mA

 

0.05%

+ 5mA

 

0.05%

+ 10mA

 

0.05%

+ 10mA

 

0.05%

+ 20mA

6V / 60V

 

24V / 240V

 

6V / 60V

 

15V / 150V

 

6V / 60V

 

 

 

 

0.05% + 3mV

 

0.05%

+ 10mV

 

0.05%

+ 3 mV

 

0.05%

+ 8mV

 

0.05%

+ 3mV

0.05% + 8mV

 

0.05%

+ 20mV

 

0.05%

+ 8 mV

 

0.05%

+ 16mV

 

0.05%

+ 8mV

0.1% + 0.5W

 

0.1% + 1.2W

 

0.1% + 0.5W

 

0.1% + 1.5W

 

0.1% + 1.2W

 

 

 

 

2Specification is +/-(% of reading + fixed offset). Measurement is 1000 samples. Specification may degrade when the unit is subject to an RF field of 3V/meter, the unit is subject to line spikes of 500V, or an 8kV electrostatic discharge.

3For example, if the duty cycle setting is 20%, the actual duty cycle would be in the range of 19% to 21%.

4DC current accuracy specifications apply 30 seconds after input current is applied.

72

Image 69
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols Declaration of Conformity DeclarationPrinting History Acoustic Noise InformationTable of Contents TURN-ON Checkout 101 Front Panel -At a Glance Quick ReferenceInstrument Configuration Rear Panel At a GlanceQuick Reference Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance Voltage Quick Reference Front Panel MenusComplete description of all programming commands Scpi Programming Commands At a GlanceQuick Reference Scpi Commands Task Where to find information Document OrientationGetting Started Map Safety Considerations Options and AccessoriesDescription Features and Capabilities Front Panel ControlsRemote Programming Constant Current CC Mode Operating ModesImmediate Current Level General Information Triggered Current Level Constant Resistance CR ModeTransient Current Level Software Current LimitConstant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Transient Voltage Level Triggered Voltage LevelContinuous Transient Operation Transient OperationPulsed Transient Operation Toggled Transient OperationList Operation Triggering a preset levelTriggered Operation Triggering a transient pulseRisetime Transition Limitation Slew Rate and Minimum Transition TimeTransition Time Slew Rate Examples Resetting Latched Protection Protection FeaturesInput Control Short On/OffOvertemperature OverpowerGeneral Information Overvoltage OvercurrentReading Remote Programming Errors Reverse VoltageStatus Reporting External Control Signals Saving and Recalling SettingsRemote Sensing Monitor OutputsFault Input MeasurementsPort On/Off DC Measurements Power MeasurementsRMS Measurements Minimum and Maximum MeasurementsInspection CleaningDamage Packaging Material Items SuppliedInstalling the Modules ProcedureInstallation Channel Number Location3A. N3300A Outline Diagram Bench Operation Rack MountingDo not block the fan exhaust at the rear of the unit Power Cord Manually-Tightened ConnectorsInput Connections 6a. Manual Connector 8mm Screw Terminal Connector option UJ1Wire Considerations Stranded Copper Wire Ampere Capacity Wire Size AmpacityMaximum Wire Lengths to Limit Voltage Drops A1-A3 Control ConnectorA11 and A10 +S and -STrigger and Digital Connections Sense SwitchComputer Connections Gpib InterfaceRS-232 Interface Local Sense Connections Application ConnectionsRemote Sense Connections Parallel Connections11. Remote Sensing 13. Zero Volt Loading Low Voltage OperationCheckout Procedure IntroductionCheckout Programming Values Selftest Errors Error MessagesSelftest Error Codes Case of TroubleFront Panel, Overall View Front Panel DescriptionFront Panel Operation System Keys Display Command FunctionImmediate Action Keys Function keysScrolling Keys Metering KeysDisplay Measurement Inpshor Input Control KeysTransient Control Keys Trigger Control KeysList Control Keys Entry Keys Entry KeysSet the constant current mode input Using the Front Panel DisplayExamples of Front Panel Programming Programming Constant Current, Voltage and Resistance ModesSet the constant voltage mode input Set the constant resistance mode inputProgramming Transient Operation Set transient operation for Continuous modeSet transient operation for Toggled mode Set transient operation for Pulse modeTmode Togg Programming ListsList Sequence Query and Clear Errors as follows Querying and Clearing Output Protection and ErrorsMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsStoring and Recalling Instrument States Setting the Gpib AddressSet the Gpib address as follows Save an instrument state in location 0 as followsSpecifications Specifications Table A-2. Supplemental Characteristics Specifications aN3302A N3303A N3304A N3305A N3306A SpecificationsEquipment Required Performance Test and Calibration ProceduresPerformance Tests CC Mode TestsImon Zero Verification Performance Test and CalibrationCC Mode Regulation Test CV Mode TestsPerformance Test and Calibration B Voltage Programming and Measurement Accuracy TestCV Mode Regulation Test CR Mode Tests Resistance Range 1 Programming AccuracyResistance Range 2 Programming Accuracy Resistance Range 4 Programming Accuracy Resistance Range 3 Programming AccuracyFrequency and Duty Cycle Accuracy Transient Generator Mode TestsOutput 705TRAN on Output 705TRIGIMM Performance Test and Calibration Pulse Width AccuracyAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Table B-2 Variables and Power Source Settings CalibrationParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Print TABXY9,15CALIBRATING High Voltage Range Voltage Calibration ProgramPerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 101 Index102 Index103 Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates