Agilent Technologies n3306a, n3305a, n3304a, n3301a, n3303a, n3302a, n3300a manual Calibrating Range

Page 95

B - Performance Test and Calibration

550PRINT TABXY(10,12);"Set power source to voltage and current settings from Table B-2"

560 PRINT TABXY(10,13);"for resistance range 1 - Point P2"

570DISP "Press Cont to continue resistance range 1 calibration"

580PAUSE

590OUTPUT @Ld;"CAL:LEV P2"

600CLEAR SCREEN

610PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 1 - P2 - 30 SECOND WAIT”

620WAIT 30

630INPUT "Measure voltage across current shunt - enter value in volts",Vp2

640Ip2=Vp2/Rshunt1

650INPUT "Measure voltage across +/- sense terminals and enter value in volts",Vt2

660Rp2=Vt2/Ip2

670OUTPUT @Ld;"CAL:DATA";Rp2

680OUTPUT @Ld;"INP:STAT OFF"

690!

700

!

********** CALIBRATING RANGE 2

******************

710

!

 

 

720

CLEAR SCREEN

 

730INPUT "Enter value of current shunt for range 2 in OHMS",Rshunt2

740PRINT TABXY(10,10);"You have entered ";Rshunt2;" ohms"

750INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$

760IF UPC$(Y$)="Y" OR Y$="" THEN 780

770IF UPC$(Y$)="N" OR Y$="n" THEN 720

780CLEAR SCREEN

790!

800PRINT TABXY(10,12);"Set power source to voltage and current settings from"

810PRINT TABXY(10,13);"Table B-1 for resistance range 2 - Point P1"

820DISP "Press CONT when ready to calibrate resistance range 2"

830PAUSE

840CLEAR SCREEN

850PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 2 - P1 - 15 SECOND WAIT"

860OUTPUT @Ld;"INP:STAT ON"

870OUTPUT @Ld;"RES:RANG 20" !< resistance range 2 variable from table B-1

880OUTPUT @Ld;"CAL:LEV P1"

890WAIT 15

900INPUT "Measure voltage across current shunt - enter value in volts",Vp1

910Ip1=Vp1/Rshunt2

920INPUT "Measure voltage across +/- sense terminals and enter value in volts",Vt1

930Rp1=Vt1/Ip1

940OUTPUT @Ld;"CAL:DATA";Rp1

950!

960CLEAR SCREEN

970PRINT TABXY(10,12);"Set power source to voltage and current settings from Table B-2"

980PRINT TABXY(10,13);"for resistance range 2 - Point P2"

990DISP "Press Cont to continue resistance range 2 calibration"

1000 PAUSE

1010 OUTPUT @Ld;"CAL:LEV P2"

1020 CLEAR SCREEN

1030 PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 2 - P2 - 15 SECOND WAIT" 1040 WAIT 15

1050 INPUT "Measure voltage across current shunt - enter value in volts",Vp2 1060 Ip2=Vp2/Rshunt2

1070 INPUT "Measure voltage across +/- sense terminals and enter value in volts",Vt2

98

Image 95
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols Declaration of Conformity DeclarationPrinting History Acoustic Noise InformationTable of Contents TURN-ON Checkout 101 Front Panel -At a Glance Quick ReferenceQuick Reference Instrument ConfigurationRear Panel At a Glance Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance Voltage Quick Reference Front Panel MenusComplete description of all programming commands Scpi Programming Commands At a GlanceQuick Reference Scpi Commands Getting Started Map Task Where to find informationDocument Orientation Description Safety ConsiderationsOptions and Accessories Remote Programming Features and CapabilitiesFront Panel Controls Immediate Current Level Constant Current CC ModeOperating Modes Software Current Limit Constant Resistance CR ModeGeneral Information Triggered Current Level Transient Current LevelTransient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Transient Voltage Level Triggered Voltage LevelToggled Transient Operation Transient OperationContinuous Transient Operation Pulsed Transient OperationTriggering a transient pulse Triggering a preset levelList Operation Triggered OperationRisetime Transition Limitation Slew Rate and Minimum Transition TimeTransition Time Slew Rate Examples Short On/Off Protection FeaturesResetting Latched Protection Input ControlOvercurrent OverpowerOvertemperature General Information OvervoltageStatus Reporting Reading Remote Programming ErrorsReverse Voltage Monitor Outputs Saving and Recalling SettingsExternal Control Signals Remote SensingPort On/Off FaultInput Measurements Minimum and Maximum Measurements Power MeasurementsDC Measurements RMS MeasurementsItems Supplied CleaningInspection Damage Packaging MaterialInstallation Installing the ModulesProcedure Channel Number Location3A. N3300A Outline Diagram Do not block the fan exhaust at the rear of the unit Bench OperationRack Mounting Input Connections Power CordManually-Tightened Connectors 6a. Manual Connector 8mm Screw Terminal Connector option UJ1Wire Considerations Maximum Wire Lengths to Limit Voltage Drops Stranded Copper Wire Ampere CapacityWire Size Ampacity +S and -S Control ConnectorA1-A3 A11 and A10Trigger and Digital Connections Sense SwitchRS-232 Interface Computer ConnectionsGpib Interface Parallel Connections Application ConnectionsLocal Sense Connections Remote Sense Connections11. Remote Sensing 13. Zero Volt Loading Low Voltage OperationCheckout Programming Values Checkout ProcedureIntroduction Case of Trouble Error MessagesSelftest Errors Selftest Error CodesFront Panel, Overall View Front Panel DescriptionFront Panel Operation System Keys Display Command FunctionImmediate Action Keys Function keysDisplay Measurement Scrolling KeysMetering Keys Inpshor Input Control KeysList Control Keys Transient Control KeysTrigger Control Keys Entry Keys Entry KeysProgramming Constant Current, Voltage and Resistance Modes Using the Front Panel DisplaySet the constant current mode input Examples of Front Panel ProgrammingSet the constant voltage mode input Set the constant resistance mode inputProgramming Transient Operation Set transient operation for Continuous modeSet transient operation for Toggled mode Set transient operation for Pulse modeTmode Togg Programming ListsList Sequence Use the Meter menu for making front panel measurements Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows Making Basic Front Panel MeasurementsSave an instrument state in location 0 as follows Setting the Gpib AddressStoring and Recalling Instrument States Set the Gpib address as followsSpecifications Specifications Table A-2. Supplemental Characteristics Specifications aN3302A N3303A N3304A N3305A N3306A SpecificationsEquipment Required Performance Test and Calibration ProceduresPerformance Test and Calibration CC Mode TestsPerformance Tests Imon Zero VerificationVoltage Programming and Measurement Accuracy Test CV Mode TestsCC Mode Regulation Test Performance Test and Calibration BCV Mode Regulation Test Resistance Range 2 Programming Accuracy CR Mode TestsResistance Range 1 Programming Accuracy Resistance Range 4 Programming Accuracy Resistance Range 3 Programming AccuracyFrequency and Duty Cycle Accuracy Transient Generator Mode TestsOutput 705TRAN on Output 705TRIGIMM Performance Test and Calibration Pulse Width AccuracyAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Parameters Calibrated Table B-2 Variables and Power Source SettingsCalibration IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Print TABXY9,15CALIBRATING High Voltage Range Voltage Calibration ProgramPerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 101 Index102 Index103 Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates