Agilent Technologies n3302a, n3305a, n3304a, n3301a, n3303a, n3306a, n3300a CV Mode Regulation Test

Page 75

B - Performance Test and Calibration

4. Turn on the power source. Set the power source voltage and current to the values listed in the following table.

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

61V

246V

61V

152V

61V

 

Power Source Current Setting

2A

0.6A

3A

2A

3A

5.

Checking high voltage range, high voltage point. Wait 10 seconds then record actual input voltage and front panel

 

 

voltage reading on test record card under high voltage range, high voltage.

 

 

 

6.

Checking high voltage range, low voltage point. Press [ VOLT ]

[ 3 ] [ Enter ]. Wait 10 seconds, then record actual

 

 

input voltage and front panel readings on test record card under high voltage range, low voltage.

 

7.

Checking low voltage range, low voltage point. Press [ VOLT ]. Scroll until display reads V:RANG Press [ 1 ]

[

 

Enter ]. Press shift key then [ Sense ] scroll till display reads S:V:RNG, press [ 3 ]. Wait 10 seconds then record actual

 

input voltage and front panel readings on test record card under low voltage range, low voltage.

 

8.

Checking low voltage range, high voltage point. Press [ VOLT ] [ select voltage from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

 

N3305A

N3306A

 

 

6V

24V

6V

 

15V

6V

 

 

9.

Wait 10 seconds then record actual input voltage and front panel readings on test record card under low voltage range,

 

low voltage.

 

 

 

 

 

 

 

CV Mode Regulation Test

This test verifies the Input Voltage remains within specification when the input current is changed from a low current to rated current.

1.

Connect Electronic Load, power source and DVM as shown in figure B-3. Be sure + sense and – sense are connected

 

to the +/- input terminals and the remote/local button is in the remote position.

 

 

 

2.

Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].

3.

Press [ Func ]. Scroll until display reads FUNC VOLT, then press [ Enter ].

 

 

 

 

 

4.

Press [ Volt ] [ select voltage from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

N3305A

 

N3306A

 

 

 

 

5V

25V

 

5V

 

8.3V

 

5V

 

 

 

5.

Turn on power source. Set power source voltage and current to values listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

6V

 

30V

 

6.5V

10V

6.6V

 

 

Power Source Current Setting

1A

 

0.4A

 

1A

1A

1A

 

6.

Wait 10 seconds and record input voltage reading.

 

 

 

 

 

 

 

 

7.

Reset power source current level to value listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Current Setting

30A

 

10A

 

60A

60A

120A

 

8.

Wait 10 seconds and record input voltage reading.

 

 

 

 

 

 

 

 

9.

Subtract reading in step ‘8’ from reading in step ‘6’. Absolute value of difference should be less then specification.

78

Image 75
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols Declaration of Conformity DeclarationPrinting History Acoustic Noise InformationTable of Contents TURN-ON Checkout 101 Front Panel -At a Glance Quick ReferenceInstrument Configuration Rear Panel At a GlanceQuick Reference Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance Voltage Quick Reference Front Panel MenusComplete description of all programming commands Scpi Programming Commands At a GlanceQuick Reference Scpi Commands Task Where to find information Document OrientationGetting Started Map Safety Considerations Options and AccessoriesDescription Features and Capabilities Front Panel ControlsRemote Programming Constant Current CC Mode Operating ModesImmediate Current Level Software Current Limit Constant Resistance CR ModeGeneral Information Triggered Current Level Transient Current LevelConstant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Transient Voltage Level Triggered Voltage LevelToggled Transient Operation Transient OperationContinuous Transient Operation Pulsed Transient OperationTriggering a transient pulse Triggering a preset levelList Operation Triggered OperationRisetime Transition Limitation Slew Rate and Minimum Transition TimeTransition Time Slew Rate Examples Short On/Off Protection FeaturesResetting Latched Protection Input ControlOvercurrent OverpowerOvertemperature General Information OvervoltageReading Remote Programming Errors Reverse VoltageStatus Reporting Monitor Outputs Saving and Recalling SettingsExternal Control Signals Remote SensingFault Input MeasurementsPort On/Off Minimum and Maximum Measurements Power MeasurementsDC Measurements RMS MeasurementsItems Supplied CleaningInspection Damage Packaging MaterialInstalling the Modules ProcedureInstallation Channel Number Location3A. N3300A Outline Diagram Bench Operation Rack MountingDo not block the fan exhaust at the rear of the unit Power Cord Manually-Tightened ConnectorsInput Connections 6a. Manual Connector 8mm Screw Terminal Connector option UJ1Wire Considerations Stranded Copper Wire Ampere Capacity Wire Size AmpacityMaximum Wire Lengths to Limit Voltage Drops +S and -S Control ConnectorA1-A3 A11 and A10Trigger and Digital Connections Sense SwitchComputer Connections Gpib InterfaceRS-232 Interface Parallel Connections Application ConnectionsLocal Sense Connections Remote Sense Connections11. Remote Sensing 13. Zero Volt Loading Low Voltage OperationCheckout Procedure IntroductionCheckout Programming Values Case of Trouble Error MessagesSelftest Errors Selftest Error CodesFront Panel, Overall View Front Panel DescriptionFront Panel Operation System Keys Display Command FunctionImmediate Action Keys Function keysScrolling Keys Metering KeysDisplay Measurement Inpshor Input Control KeysTransient Control Keys Trigger Control KeysList Control Keys Entry Keys Entry KeysProgramming Constant Current, Voltage and Resistance Modes Using the Front Panel DisplaySet the constant current mode input Examples of Front Panel ProgrammingSet the constant voltage mode input Set the constant resistance mode inputProgramming Transient Operation Set transient operation for Continuous modeSet transient operation for Toggled mode Set transient operation for Pulse modeTmode Togg Programming ListsList Sequence Use the Meter menu for making front panel measurements Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows Making Basic Front Panel MeasurementsSave an instrument state in location 0 as follows Setting the Gpib AddressStoring and Recalling Instrument States Set the Gpib address as followsSpecifications Specifications Table A-2. Supplemental Characteristics Specifications aN3302A N3303A N3304A N3305A N3306A SpecificationsEquipment Required Performance Test and Calibration ProceduresPerformance Test and Calibration CC Mode TestsPerformance Tests Imon Zero VerificationVoltage Programming and Measurement Accuracy Test CV Mode TestsCC Mode Regulation Test Performance Test and Calibration BCV Mode Regulation Test CR Mode Tests Resistance Range 1 Programming AccuracyResistance Range 2 Programming Accuracy Resistance Range 4 Programming Accuracy Resistance Range 3 Programming AccuracyFrequency and Duty Cycle Accuracy Transient Generator Mode TestsOutput 705TRAN on Output 705TRIGIMM Performance Test and Calibration Pulse Width AccuracyAgilent N3302A Verification Test Record Agilent N3303A Verification Test Record Agilent N3304A Verification Test Record Agilent N3305A Verification Test Record Agilent N3306A Verification Test Record Calibration Figure B-3. Voltage Calibration Table B-2 Variables and Power Source Settings CalibrationParameters Calibrated IMON, Iprog and Current Calibration Program Performance Test and Calibration Wait Performance Test and Calibration Print TABXY9,15CALIBRATING High Voltage Range Voltage Calibration ProgramPerformance Test and Calibration Resistance Calibration Program Calibrating Range Performance Test and Calibration B 100 101 Index102 Index103 Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates