Teledyne 3000ZA-3X Function Keys, Data Entry Keys, Digital Meter Display, Flowmeter

Page 16

Introduction

3000ZA-3X

 

 

 

Function Keys:

Six touch-sensitive membrane switches are used to change the specific function performed by the analyzer:

Analyze Perform analysis for oxygen content of a sample gas.

System Perform system-related tasks (described in detail in chapter 4, Operation.).

Span Span calibrate the analyzer.

Zero Zero calibrate the analyzer.

Alarms Set the alarm setpoints and attributes.

Range Set up the 3 user definable ranges for the instrument.

Data Entry Keys:

Six touch-sensitive membrane switches are used to input data to the instrument via the alphanumeric VFD display:

Left & Right Arrows Select between functions currently displayed on the VFD screen.

Up & Down Arrows Increment or decrement values of functions currently displayed.

Enter Advances VFD display to the next screen in a series or returns to the Analyze screen if none remain.

Escape Backs VFD display to the previous screen in a series or returns to the Analyze screen if none remain.

Digital Meter Display:

The meter display is a Light Emitting Diode (LED) device that produces large, bright, 7-segment numbers that are legible in any lighting. It produces a continuous readout from 0-250 ppm. It is accurate across all analysis ranges without the discontinuity inherent in analog range switching.

Alphanumeric Interface Screen:

The VFD screen is an easy-to-use interface from operator to analyzer. It displays values, options, and messages that give the operator immediate feedback.

Flowmeter:

Monitors the flow of gas past the sensor. Readout is 0.2 to 2.4 standard liters per minute (SLPM).

Teledyne Analytical Instruments

4

Image 16
Contents Model 3000ZA-3X 3000ZA-3X Specific Model Information Important Notice Safety Messages 3000ZA-3X Vii Table of Contents Testing the System Maintenance AppendixModel 3000ZA Front Panel Model 3000ZA Rear PanelList of Tables XiiMain Features of the Analyzer IntroductionOverview Typical ApplicationsModel Designations Introduction 3000ZA-3XFront Panel Operator Interface Model 3000ZA Front PanelFunction Keys Data Entry KeysDigital Meter Display Alphanumeric Interface ScreenRear Panel Equipment Interface Access DoorStandby Button ∙ 50-Pin Equipment Interface Port OptionalOperational Theory IntroductionZrO2 Sensor Principles of OperationSample System Operational Theory 3000ZA-3XElectronics and Signal Processing Zirconium Oxide Sensor Application NotesOperational Theory 3000ZA-3X Component Location Electronic Block Diagram Unpacking the Analyzer InstallationMounting the Analyzer Installation 3000ZA-3X Rear Panel Connections Gas ConnectionsInstallation3000ZA-3X Zero in and Span in Optional Electrical ConnectionsEquipment Interface Connector Pin Arrangement Threshold Alarm Analog Output ConnectionsSystem Alarm Alarm Relay Contact PinsRemote Calibration Connections Range ID Relay Connections Remote Probe Connections FET Series Resistance 2.3 RS-232 PortCommands via RS-232 Input Installing the Zirconium Oxide Sensors Testing the SystemOperation Style Conventions Using the Data Entry and Function ButtonsKeys Description Enter EscapeAnalyze Screen Menu/Screen StructureSystem Menu Hierarchy of Available Functions Password routine Enter Password T aChange Password Enter New Password T aLogout Routine Model RoutineSelf-Test Routine Analog-Out Adjust RoutineFilter Routine 20mA Gain 20mA Limit False Track in Cal Yes Span Menu Auto-Calibration TimerChange Stream Span Sensor Span TimerSpan Value Span Finish Span Auto FinishSpan Begin Span Begin Begin Span?Zero Finish Zero MenuZero Value Zero Timer Zero Timer 7 minZero Begin Zero Begin Begin Zero?Alarms Menu Alarm ActiveAlarm Polarity AL-1 NON-LATCHING AL-1 20.00 ppm AL-2 ActiveAlarm Failsafe Alarm LatchingAlarm Setpoint AL-1 80.00 ppmRange Select Range MenuRange Select Auto Manual Range Select Range SetpointsMaintenance Routine MaintenanceSensor Replacement Removing the SensorMaintenance 3000ZA-3X Fuse Replacement Sensor Block with Middle Sensor InstalledSystem Self Diagnostic Test Self-Test Result DisplayMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingTrace Oxygen Analyzer Maintenance Maintenance 3000ZA-3X Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-35 C Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsRestriction Devices BY-PASS Conversions