Teledyne 3000ZA-3X operating instructions Remote Calibration Connections

Page 33

Trace Oxygen Analyzer

Installation

 

 

 

Zero: Floating input. 5 to 24 V input across the + and – pins puts the analyzer into the Zero mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Zero mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately. See Remote Probe Connector. (The –C option internal valves operate automatically.)

Span: Floating input. 5 to 24 V input across the + and – pins puts the analyzer into the Span mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Span mode to terminate when done. A synchronous signal must open and close external span valve appropriately. See Figure 3-5 Remote Probe Connector. (The –C option internal valves operate automatically.)

Cal Contact: This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.)

Table 3-3: Remote Calibration Connections

Pin Function

9+ Remote Zero

11 – Remote Zero

10 + Remote Span

12 – Remote Span

40 Cal Contact

41 Cal Contact

Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 3000ZA Analyzer, the customer's controller must monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.

 

 

 

Teledyne Analytical Instruments

21

Image 33
Contents Model 3000ZA-3X 3000ZA-3X Specific Model Information Important Notice Safety Messages 3000ZA-3X Vii Table of Contents Testing the System Appendix MaintenanceModel 3000ZA Rear Panel Model 3000ZA Front PanelXii List of TablesIntroduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction 3000ZA-3X Model DesignationsModel 3000ZA Front Panel Front Panel Operator InterfaceData Entry Keys Function KeysDigital Meter Display Alphanumeric Interface ScreenAccess Door Rear Panel Equipment InterfaceStandby Button Optional ∙ 50-Pin Equipment Interface PortIntroduction Operational TheoryZrO2 Sensor Principles of OperationOperational Theory 3000ZA-3X Sample SystemZirconium Oxide Sensor Application Notes Electronics and Signal ProcessingOperational Theory 3000ZA-3X Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Installation 3000ZA-3X Gas Connections Rear Panel ConnectionsInstallation3000ZA-3X Electrical Connections Zero in and Span in OptionalEquipment Interface Connector Pin Arrangement Analog Output Connections Threshold AlarmAlarm Relay Contact Pins System AlarmRemote Calibration Connections Range ID Relay Connections Remote Probe Connections 2.3 RS-232 Port FET Series ResistanceCommands via RS-232 Input Testing the System Installing the Zirconium Oxide SensorsOperation Using the Data Entry and Function Buttons Style ConventionsKeys Description Escape EnterMenu/Screen Structure Analyze ScreenSystem Menu Hierarchy of Available Functions Enter Password T a Password routineChange Password Enter New Password T aModel Routine Logout RoutineAnalog-Out Adjust Routine Self-Test RoutineFilter Routine 20mA Gain 20mA Limit False Track in Cal Yes Auto-Calibration Timer Span MenuChange Stream Span Timer Span SensorSpan Value Span Auto Finish Span FinishSpan Begin Span Begin Begin Span?Zero Menu Zero FinishZero Value Zero Timer 7 min Zero TimerZero Begin Zero Begin Begin Zero?Alarm Active Alarms MenuAlarm Polarity AL-1 NON-LATCHING AL-1 20.00 ppm AL-2 ActiveAlarm Latching Alarm FailsafeAlarm Setpoint AL-1 80.00 ppmRange Menu Range SelectRange Select Auto Range Setpoints Manual Range SelectRoutine Maintenance MaintenanceSensor Replacement Removing the SensorMaintenance 3000ZA-3X Sensor Block with Middle Sensor Installed Fuse ReplacementSelf-Test Result Display System Self Diagnostic TestRear Panel Removal Remove only those screws marked with an Major Internal ComponentsTroubleshooting CleaningTrace Oxygen Analyzer Maintenance Maintenance 3000ZA-3X Teledyne Analytical Instruments Appendix SpecificationsOperating Temperature 5-35 C Qty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListFlow Rate Recommendations Application NotesRestriction Devices BY-PASS Conversions