Maintenance |
| |
|
|
|
2.Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool.
3.Remove the eight screws that hold the cover of the sensor block. As you remove the cover make sure the
4.You can remove the sensor that you want to replace with your fingers or using small pliers. Just pull the sensor straight out.
CAUTION: SENSORS MIGHT BE HOT! AFTER REMOVING POWER. LET THE SENSORS COOL FOR A FEW MINUTES.
5.Install the new sensors paying attention to relative position in the sensor block. Note orientation of sensor tab aligning with tab location on the silkscreen of the mounting PCB. Sensor is held by friction of the pins sockets. The locations of the Sensors are:
On the left side P/N S1439
On the right side P/N S1442 (0 – 95 % range )
CAUTION: MAKE SURE OF THE CORRECT POSITION OF EACH SENSOR. DO NOT PLACE A SENSOR IN THE WRONG SOCKET OR THE ANALYZER WILL FAIL!
Teledyne Analytical Instruments | 46 |