Page 58

Maintenance

3000ZA-3X

 

 

 

2.Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool.

3.Remove the eight screws that hold the cover of the sensor block. As you remove the cover make sure the O-rings stay in place. If they drop, place them back carefully in the groove found on the cell block, not on the cover. See Figure 5.1.

4.You can remove the sensor that you want to replace with your fingers or using small pliers. Just pull the sensor straight out.

CAUTION: SENSORS MIGHT BE HOT! AFTER REMOVING POWER. LET THE SENSORS COOL FOR A FEW MINUTES.

5.Install the new sensors paying attention to relative position in the sensor block. Note orientation of sensor tab aligning with tab location on the silkscreen of the mounting PCB. Sensor is held by friction of the pins sockets. The locations of the Sensors are:

On the left side P/N S1439 (0-200 ppm range ) On the center P/N S1492 (0 – 2 % range )

On the right side P/N S1442 (0 – 95 % range )

CAUTION: MAKE SURE OF THE CORRECT POSITION OF EACH SENSOR. DO NOT PLACE A SENSOR IN THE WRONG SOCKET OR THE ANALYZER WILL FAIL!

Teledyne Analytical Instruments

46

Image 58
Contents Model 3000ZA-3X 3000ZA-3X Specific Model Information Important Notice Safety Messages 3000ZA-3X Vii Table of Contents Testing the System Maintenance AppendixModel 3000ZA Front Panel Model 3000ZA Rear PanelList of Tables XiiOverview Main Features of the AnalyzerIntroduction Typical ApplicationsModel Designations Introduction 3000ZA-3XFront Panel Operator Interface Model 3000ZA Front PanelDigital Meter Display Function KeysData Entry Keys Alphanumeric Interface ScreenRear Panel Equipment Interface Access DoorStandby Button ∙ 50-Pin Equipment Interface Port OptionalZrO2 Sensor Operational TheoryIntroduction Principles of OperationSample System Operational Theory 3000ZA-3XElectronics and Signal Processing Zirconium Oxide Sensor Application NotesOperational Theory 3000ZA-3X Component Location Electronic Block Diagram Unpacking the Analyzer InstallationMounting the Analyzer Installation 3000ZA-3X Rear Panel Connections Gas ConnectionsInstallation3000ZA-3X Zero in and Span in Optional Electrical ConnectionsEquipment Interface Connector Pin Arrangement Threshold Alarm Analog Output ConnectionsSystem Alarm Alarm Relay Contact PinsRemote Calibration Connections Range ID Relay Connections Remote Probe Connections FET Series Resistance 2.3 RS-232 PortCommands via RS-232 Input Installing the Zirconium Oxide Sensors Testing the SystemOperation Style Conventions Using the Data Entry and Function ButtonsKeys Description Enter EscapeAnalyze Screen Menu/Screen StructureSystem Menu Hierarchy of Available Functions Change Password Password routineEnter Password T a Enter New Password T aLogout Routine Model RoutineSelf-Test Routine Analog-Out Adjust RoutineFilter Routine 20mA Gain 20mA Limit False Track in Cal Yes Span Menu Auto-Calibration TimerChange Stream Span Sensor Span TimerSpan Value Span Begin Span FinishSpan Auto Finish Span Begin Begin Span?Zero Finish Zero MenuZero Value Zero Begin Zero TimerZero Timer 7 min Zero Begin Begin Zero?Alarm Polarity Alarms MenuAlarm Active AL-1 NON-LATCHING AL-1 20.00 ppm AL-2 ActiveAlarm Setpoint Alarm FailsafeAlarm Latching AL-1 80.00 ppm Range Select Range Menu Range Select Auto Manual Range Select Range SetpointsSensor Replacement MaintenanceRoutine Maintenance Removing the SensorMaintenance 3000ZA-3X Fuse Replacement Sensor Block with Middle Sensor InstalledSystem Self Diagnostic Test Self-Test Result DisplayMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingTrace Oxygen Analyzer Maintenance Maintenance 3000ZA-3X Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-35 C Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsRestriction Devices BY-PASS Conversions