Trace Oxygen Analyzer | Maintenance | |
|
|
|
Maintenance
5.1 Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing the sensors and fuses, and recalibration. For recalibration, see Section 4.4 Calibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
5.2 Sensor Replacement
The ZrO2 sensors are sealed transducers with no electrolyte to change or electrodes to clean. When the sensor reaches the end of its useful life, it is replaced. The spent sensor should be discarded according to local regulations.
5.2.1 Removing the Sensor
The ZrO2 sensors is located inside the nylon cell block behind the front panel.
To remove an existing cell (refer to Figure
CAUTION: POWER IS APPLIED TO THE INSTRUMENT CIRCUITRY WHENEVER THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. THE RED SWITCH ON THE FRONT PANEL IS FOR SWITCHING POWER ON OR OFF TO THE DISPLAYS AND OUTPUTS ONLY.
1.Remove power to the instrument by unplugging the power cord at the power source.
|
|
|
Teledyne Analytical Instruments | 45 |