Teledyne 3000ZA-3X operating instructions Equipment Interface Connector Pin Arrangement

Page 30

Installation

3000ZA-3X

 

 

 

CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. THE RED SWITCH ON THE FRONT PANEL IS FOR SWITCHING POWER ON OR OFF TO THE DISPLAYS AND OUTPUTS ONLY.

The universal power supply requires an 85–250 VAC, 47-63 Hz power source.

Fuse Installation: The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation. (See Fuse Replacement in Chapter 5, Maintenance.)

3.3.2.250-PIN EQUIPMENT INTERFACE CONNECTOR

Figure 3-4 shows the pin layout of the Equipment Interface Connector. The arrangement is shown as seen when the viewer faces the rear panel of the analyzer. The pin numbers for each input/output function are given where each function is described in the paragraphs below.

Figure 3-4: Equipment Interface Connector Pin Arrangement

Analog Outputs: There are four DC output signal pins—two pins per output. For polarity, see Table 3-1. The outputs are:

0–1 VDC % of Range:Voltage rises linearly with increasing oxygen, from 0 V at 0 ppm to 1 V at full scale ppm. (Full scale = 100% of programmable range.)

0–1 VDC Range ID: 0.20 V = Low Range, 0.5 V = Medium Range, 0.80 V = High Range.

Teledyne Analytical Instruments

18

Image 30
Contents Model 3000ZA-3X 3000ZA-3X Specific Model Information Important Notice Safety Messages 3000ZA-3X Vii Table of Contents Testing the System Maintenance AppendixModel 3000ZA Front Panel Model 3000ZA Rear PanelList of Tables XiiOverview Main Features of the AnalyzerIntroduction Typical ApplicationsModel Designations Introduction 3000ZA-3XFront Panel Operator Interface Model 3000ZA Front PanelDigital Meter Display Function KeysData Entry Keys Alphanumeric Interface ScreenAccess Door Rear Panel Equipment InterfaceStandby Button ∙ 50-Pin Equipment Interface Port OptionalZrO2 Sensor Operational TheoryIntroduction Principles of OperationSample System Operational Theory 3000ZA-3XElectronics and Signal Processing Zirconium Oxide Sensor Application NotesOperational Theory 3000ZA-3X Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Installation 3000ZA-3X Rear Panel Connections Gas ConnectionsInstallation3000ZA-3X Zero in and Span in Optional Electrical ConnectionsEquipment Interface Connector Pin Arrangement Threshold Alarm Analog Output ConnectionsSystem Alarm Alarm Relay Contact PinsRemote Calibration Connections Range ID Relay Connections Remote Probe Connections FET Series Resistance 2.3 RS-232 PortCommands via RS-232 Input Installing the Zirconium Oxide Sensors Testing the SystemOperation Using the Data Entry and Function Buttons Style ConventionsKeys Description Enter EscapeMenu/Screen Structure Analyze ScreenSystem Menu Hierarchy of Available Functions Change Password Password routineEnter Password T a Enter New Password T aLogout Routine Model RoutineAnalog-Out Adjust Routine Self-Test RoutineFilter Routine 20mA Gain 20mA Limit False Track in Cal Yes Auto-Calibration Timer Span MenuChange Stream Span Timer Span SensorSpan Value Span Begin Span FinishSpan Auto Finish Span Begin Begin Span?Zero Menu Zero FinishZero Value Zero Begin Zero TimerZero Timer 7 min Zero Begin Begin Zero?Alarm Polarity Alarms MenuAlarm Active AL-1 NON-LATCHING AL-1 20.00 ppm AL-2 ActiveAlarm Setpoint Alarm FailsafeAlarm Latching AL-1 80.00 ppmRange Menu Range SelectRange Select Auto Manual Range Select Range SetpointsSensor Replacement MaintenanceRoutine Maintenance Removing the SensorMaintenance 3000ZA-3X Fuse Replacement Sensor Block with Middle Sensor InstalledSystem Self Diagnostic Test Self-Test Result DisplayMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingTrace Oxygen Analyzer Maintenance Maintenance 3000ZA-3X Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-35 C Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsRestriction Devices BY-PASS Conversions