Operational Theory |
| |
|
|
|
One of the three sensors has been modified
2.3 Sample System
The sample system delivers gases to the ZrO2 sensors, located on the cell block, from the analyzer rear panel inlet. Depending on the mode of operation either sample or calibration gas is delivered.
The Model 3000ZA sample system is designed and fabricated to ensure that the oxygen concentration of the gas is not altered as it travels through the sample system.
The sample system for the standard instrument incorporates compression tube fittings for sample inlet and 1/4” outlet tube connections at the rear panel. The sample or calibration gas that flows through the system is monitored by a flowmeter downstream from the cell. Figure
Figure 2-2 shows the flow diagram for sampling systems used for the three configurations offered as options for this instrument. In the standard instrument, calibration gases can be connected directly to the Sample In port by teeing to the port with appropriate valves.
Figure 2-1: Model 3000ZA-3X Sample System
Teledyne Analytical Instruments | 8 |