Teledyne 3000ZA operating instructions Remote Calibration Connections

Page 34

Installation

3000ZA

 

 

 

Span: Floating input. 5 to 24 V input across the + and – pins puts the analyzer into the Span mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Span mode to terminate when done. A synchronous signal must open and close external span valve appropriately. See Figure 3-5 Remote Probe Connector. (The –C option internal valves operate automatically.)

Cal Contact: This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.)

Table 3-3: Remote Calibration Connections

Pin Function

9+ Remote Zero

11 – Remote Zero

10 + Remote Span

12 – Remote Span

40 Cal Contact

41 Cal Contact

Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 3000ZA Analyzer, the customer's controller must monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.

For example:

1.Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.)

2.When the CRC closes, remove the zero command.

Teledyne Analytical Instruments

22

Image 34
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Maintenance OperationAppendix IndexList of Figures Model 3000ZA Front PanelList of Tables Overview Main Features of the AnalyzerIntroduction Typical ApplicationsModel Designations Introduction 3000ZAFront Panel Operator Interface Model 3000ZA Front PanelIntroduction 3000ZA Rear Panel Equipment Interface Model 3000ZA Rear PanelPower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day ZrO2 Sensor Operational TheoryIntroduction Principles of OperationSample System Model 3000ZA Sample SystemElectronics and Signal Processing Zirconium Oxide Sensor Application NotesTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Unpacking the Analyzer InstallationMounting the Analyzer Trace Oxygen Analyzer Installation Rear Panel Connections Gas ConnectionsSample Primary Input Power Electrical ConnectionsEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections Commands via RS-232 Input 2.3 RS-232 PortInstalling the Zirconium Oxide Sensor Testing the SystemTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify System Self-Diagnostic Test LogoutVersion Screen Showing Negative Oxygen ReadingsZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionFixed Range Analysis Signal Output Analyze FunctionRange ID Output Voltage Page Cell Replacement MaintenanceRoutine Maintenance Removing the SensorFuse Replacement Trace Oxygen Analyzer MaintenanceSystem Self Diagnostic Test Installing FusesMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingIf erratic behavior Specifications AppendixTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments