Teledyne 3000ZA operating instructions Alarm Relay Contact Pins

Page 33

Trace Oxygen Analyzer

Installation

 

 

 

Actuates when DC power supplied to circuits is unacceptable in one or more parameters. Permanently configured as failsafe and latching. Cannot be defeated. Actuates if self test fails.

(Reset by pressing button to remove power. Then press again and any other button EXCEPT System to resume.

Further detail can be found in Chapter 4, Section 4-5.

Table 3-2: Alarm Relay Contact Pins

Pin Contact

45Threshold Alarm 1, normally closed contact

28Threshold Alarm 1, moving contact

46Threshold Alarm 1, normally open contact

42Threshold Alarm 2, normally closed contact

44Threshold Alarm 2, moving contact

43Threshold Alarm 2, normally open contact

36System Alarm, normally closed contact

20System Alarm, moving contact

37System Alarm, normally open contact

Digital Remote Cal Inputs: Accept 0 V (off) or 24 VDC (on) inputs for remote control of calibration. (See Remote Calibration Protocol below.) See Table 3-3 for pin connections.

Zero: Floating input. 5 to 24 V input across the + and – pins puts the analyzer into the Zero mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Zero mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately. See Remote Probe Connector. (The –C option internal valves operate automatically.)

Teledyne Analytical Instruments

21

Image 33
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Operation MaintenanceIndex AppendixModel 3000ZA Front Panel List of FiguresList of Tables Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction 3000ZA Model DesignationsModel 3000ZA Front Panel Front Panel Operator InterfaceIntroduction 3000ZA Model 3000ZA Rear Panel Rear Panel Equipment InterfacePower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day Introduction Operational TheoryZrO2 Sensor Principles of OperationModel 3000ZA Sample System Sample SystemZirconium Oxide Sensor Application Notes Electronics and Signal ProcessingTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Trace Oxygen Analyzer Installation Gas Connections Rear Panel ConnectionsSample Electrical Connections Primary Input PowerEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections 2.3 RS-232 Port Commands via RS-232 InputTesting the System Installing the Zirconium Oxide SensorTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify Logout System Self-Diagnostic TestShowing Negative Oxygen Readings Version ScreenZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Analyze Function Signal OutputRange ID Output Voltage Page Routine Maintenance MaintenanceCell Replacement Removing the SensorTrace Oxygen Analyzer Maintenance Fuse ReplacementInstalling Fuses System Self Diagnostic TestRear Panel Removal Remove only those screws marked with an Major Internal ComponentsTroubleshooting CleaningIf erratic behavior Appendix SpecificationsTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing ListFlow Rate Recommendations Application NotesBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments