Teledyne 3000ZA operating instructions Span Failure

Page 56

Operation

3000ZA

 

 

 

Press Enter to enter the span value into the system and begin the span calibration.

Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as a Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates the rate of change of the Span reading. It is a sensitive indicator of stability.

####% Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1% or less of the full scale of the range being calibrated, for a period of ten minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value. The instrument then automatically enters the Analyze function.

4.4.3 Span Failure

The analyzer checks the output of the cell at the end of the span. If the raw output of the cell is too low, the span will not be accepted. The analyzer will return to the previous calibration values, trigger the System Alarm, and display in the VFD:

Span Failed!!

This message will be shown for five seconds and the instrument shall return to the Analyze mode. In the upper right hand corner of the VFD display “FCAL” will be shown. This message flag will help the operator troubleshoot in case calibration was initiated remotely. To reset the alarm and the flag message, the unit must be turned off by cycling the standby key . It will not reset if the next span cycle is correct.

A trace cell is unlikely to fail span. As explained before, when the sensor reaches the end of its useful life, the zero offset begins to rise until the analyzer finds the zero unsatisfactory. Nevertheless, feeding the wrong span gas or electronics failure could set this feature off at the end of the span. Consider this before replacing the cell.

Teledyne Analytical Instruments

44

Image 56
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Maintenance OperationAppendix IndexList of Figures Model 3000ZA Front PanelList of Tables Main Features of the Analyzer IntroductionOverview Typical ApplicationsModel Designations Introduction 3000ZAFront Panel Operator Interface Model 3000ZA Front PanelIntroduction 3000ZA Rear Panel Equipment Interface Model 3000ZA Rear PanelPower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day Operational Theory IntroductionZrO2 Sensor Principles of OperationSample System Model 3000ZA Sample SystemElectronics and Signal Processing Zirconium Oxide Sensor Application NotesTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Mounting the Analyzer InstallationUnpacking the Analyzer Trace Oxygen Analyzer Installation Rear Panel Connections Gas ConnectionsSample Primary Input Power Electrical ConnectionsEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections Commands via RS-232 Input 2.3 RS-232 PortInstalling the Zirconium Oxide Sensor Testing the SystemTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify System Self-Diagnostic Test LogoutVersion Screen Showing Negative Oxygen ReadingsZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionFixed Range Analysis Signal Output Analyze FunctionRange ID Output Voltage Page Maintenance Routine MaintenanceCell Replacement Removing the SensorFuse Replacement Trace Oxygen Analyzer MaintenanceSystem Self Diagnostic Test Installing FusesMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingIf erratic behavior Specifications AppendixTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments