Trace Oxygen Analyzer | Operation | |
|
|
|
4.3 The System Function
The sub fuctions of the System function are described below. Specific procedures for their use follow the descriptions:
•
•PSWD: Security can be established by choosing a 5 digit password (PSWD) from the standard ASCII character set. (See Installing or Changing the Password, below, for a table of ASCII characters available.) Once a unique password is assigned and activated, the operator MUST enter the UNIQUE password to gain access to
•After a password is assigned, the operator must log out to activate it. Until then, anyone can continue to operate the instrument without entering the new password.
•Only one password can be defined. Before a unique password is assigned, the system assigns TETAI by default. This allows access to anyone. After a unique password is assigned, to defeat the security, the password must be changed back to TETAI.
•Logout: Logging out prevents unauthorized tampering with analyzer settings.
•More: Select and enter More to get a new screen with additional sub functions listed.
•
•Version: Displays Manufacturer, Model, and Software Version of instrument.
•Show Negative: The operator selects whether display can show negative oxygen readings or not.
•TRAK/HLD: The operator sets whether the instrument analog outputs track the concentration change during calibration and sets a time delay for the concentration alarms after calibration
Teledyne Analytical Instruments | 31 |