Teledyne 3000ZA operating instructions Alarms Function

Page 57

Trace Oxygen Analyzer

Operation

 

 

 

4.5 The Alarms Function

The Model 3000ZA is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a set of form “C" contacts rated for 3 amperes resistive load at 250 VAC. See Table 3-2 in Chapter 3, Installation and/or the Interconnection Diagram included at the back of this manual for relay terminal connections.

The system failure alarm has a fixed configuration as described in chapter 3 Installation.

The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as latching or non-latching, and either failsafe or non-failsafe, or, they can be defeated altogether. The setpoints for the alarms are also established using this function.

Decide how your alarms should be configured. The choice will depend upon your process. Consider the following four points:

1.Which if any of the alarms are to be high alarms and which if any are to be low alarms?

2.Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint. Setting an alarm as LOW triggers the alarm when the oxygen concentration falls below the setpoint.

3.Decide whether you want the alarms to be set as:

Both high (high and high-high) alarms, or

One high and one low alarm, or

Both low (low and low-low) alarms.

4.Are either or both of the alarms to be configured as failsafe?

5.In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non-failsafe mode.

6.Are either of the alarms to be latching?

7.In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert

Teledyne Analytical Instruments

45

Image 57
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Operation MaintenanceIndex AppendixModel 3000ZA Front Panel List of FiguresList of Tables Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction 3000ZA Model DesignationsModel 3000ZA Front Panel Front Panel Operator InterfaceIntroduction 3000ZA Model 3000ZA Rear Panel Rear Panel Equipment InterfacePower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day Introduction Operational TheoryZrO2 Sensor Principles of OperationModel 3000ZA Sample System Sample SystemZirconium Oxide Sensor Application Notes Electronics and Signal ProcessingTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Trace Oxygen Analyzer Installation Gas Connections Rear Panel ConnectionsSample Electrical Connections Primary Input PowerEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections 2.3 RS-232 Port Commands via RS-232 InputTesting the System Installing the Zirconium Oxide SensorTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify Logout System Self-Diagnostic TestShowing Negative Oxygen Readings Version ScreenZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Analyze Function Signal OutputRange ID Output Voltage Page Routine Maintenance MaintenanceCell Replacement Removing the SensorTrace Oxygen Analyzer Maintenance Fuse ReplacementInstalling Fuses System Self Diagnostic TestRear Panel Removal Remove only those screws marked with an Major Internal ComponentsTroubleshooting CleaningIf erratic behavior Appendix SpecificationsTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing ListFlow Rate Recommendations Application NotesBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments