Teledyne 3000ZA operating instructions Zero and Span Functions

Page 51

Trace Oxygen Analyzer

Operation

 

 

 

TRAK/HLD Auto—Cal

PSWD Logout More

Use the __ arrow keys to blink More, then press Enter.

Version Self—Test

Show_Negative=NO

Use the __ arrow keys and select “ Show_Negative = NO”. Use the __ arrow keys to toggle from NO to YES.

Press the Escape key twice to return to the analyze mode.

This preference is stored in non-volatile memory so the configuration is remembered after a power shutdown. If the instrument is cold started, it will go back to the default case where negative oxygen readings are not shown.

4.4 The Zero and Span Functions

Zeroing is not required in order to achieve the published accuracy specification of this unit however, zeroing will eliminate offset error contributed by the sensor, electronics and the internal and external sampling systems. It will improve performance beyond published specification limits.

The analyzer is calibrated using zero and span gases.

Any suitable oxygen-free gas can be used for zero gas as long as it is known to be oxygen free and does not react adversely with the sample system.

A span gas with a known oxygen concentration in the range of 70–100% of full scale of the range of interest is recommended. In the 0- 10 ppm range, 8 ppm span gas would be best suited.

Connect the calibration gases to the analyzer according to the instructions given in Section 3.3.1, Gas Connections, observing all the prescribed precautions.

Shut off the gas pressure before connecting it to the analyzer, and be sure to limit the pressure to 40 psig or less when turning it back on.

Readjust the gas pressure into the analyzer until the flowrate (as read on the analyzer’s SLPM flowmeter) settles between 0.1 and 2.4 SLPM (approximately 0.2 - 5 SCFH).

Teledyne Analytical Instruments

39

Image 51
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Operation MaintenanceIndex AppendixModel 3000ZA Front Panel List of FiguresList of Tables Typical Applications Main Features of the AnalyzerIntroduction OverviewIntroduction 3000ZA Model DesignationsModel 3000ZA Front Panel Front Panel Operator InterfaceIntroduction 3000ZA Model 3000ZA Rear Panel Rear Panel Equipment InterfacePower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day Principles of Operation Operational TheoryIntroduction ZrO2 SensorModel 3000ZA Sample System Sample SystemZirconium Oxide Sensor Application Notes Electronics and Signal ProcessingTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Trace Oxygen Analyzer Installation Gas Connections Rear Panel ConnectionsSample Electrical Connections Primary Input PowerEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections 2.3 RS-232 Port Commands via RS-232 InputTesting the System Installing the Zirconium Oxide SensorTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify Logout System Self-Diagnostic TestShowing Negative Oxygen Readings Version ScreenZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Analyze Function Signal OutputRange ID Output Voltage Page Removing the Sensor MaintenanceRoutine Maintenance Cell ReplacementTrace Oxygen Analyzer Maintenance Fuse ReplacementInstalling Fuses System Self Diagnostic TestRear Panel Removal Remove only those screws marked with an Major Internal ComponentsTroubleshooting CleaningIf erratic behavior Appendix SpecificationsTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing ListFlow Rate Recommendations Application NotesBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments