Teledyne 3000ZA operating instructions If erratic behavior

Page 69

Trace Oxygen Analyzer

Maintenance

 

 

 

 

 

analyzer.

 

 

2. If erratic behavior

 

 

continues, replace

 

 

sensor.

 

 

 

 

Atmospheric oxygen

Increase flow rate

 

diffusing through vent

and/or length of vent

 

and altering the

tubing to dilute or

 

oxygen level the

minimize oxygen

 

sensor sees

diffusion from vent to

 

 

sensor.

 

 

 

Inaccurate zero

 

Turn analyzer off,

operation (i.e. user has

 

then back on again.

zeroed the instrument

 

Press System key

using a gas concen-

 

when prompted by

tration much higher

 

analyzer “Press

than normal for a zero

 

System for Default

gas).

 

Values”. This will

 

 

return the analyzer to

 

 

default calibration and

 

 

zero settings. Now

 

 

proceed to carefully

 

 

calibrate and zero the

 

 

analyzer.

 

 

 

Teledyne Analytical Instruments

57

Image 69
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Operation MaintenanceIndex AppendixModel 3000ZA Front Panel List of FiguresList of Tables Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction 3000ZA Model DesignationsModel 3000ZA Front Panel Front Panel Operator InterfaceIntroduction 3000ZA Model 3000ZA Rear Panel Rear Panel Equipment InterfacePower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day Introduction Operational TheoryZrO2 Sensor Principles of OperationModel 3000ZA Sample System Sample SystemZirconium Oxide Sensor Application Notes Electronics and Signal ProcessingTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Trace Oxygen Analyzer Installation Gas Connections Rear Panel ConnectionsSample Electrical Connections Primary Input PowerEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections 2.3 RS-232 Port Commands via RS-232 InputTesting the System Installing the Zirconium Oxide SensorTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify Logout System Self-Diagnostic TestShowing Negative Oxygen Readings Version ScreenZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Analyze Function Signal OutputRange ID Output Voltage Page Routine Maintenance MaintenanceCell Replacement Removing the SensorTrace Oxygen Analyzer Maintenance Fuse ReplacementInstalling Fuses System Self Diagnostic TestRear Panel Removal Remove only those screws marked with an Major Internal ComponentsTroubleshooting CleaningIf erratic behavior Appendix SpecificationsTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing ListFlow Rate Recommendations Application NotesBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments