Teledyne 3000ZA operating instructions Span Cal

Page 54

Operation

3000ZA

 

 

 

CELL FAIL/ ZERO HIGH

Before replacing the cell:

a.Check your span gas to make sure it is within specifications.

b.Check for leaks up-stream from the cell, where oxygen may be leaking into the system.

If there are no leaks and the span gas is within specification, replace the cell as described in chapter 5, Maintenance.

4.4.2 Span Cal

The Span button on the front panel is used to span calibrate the analyzer. Span calibration can be performed using the automatic mode, where an internal algorithm compares consecutive readings from the sensor to determine when the output matches the span gas concentration. Span calibration can also be performed in manual mode, where the operator determines when the span concentration reading is acceptable and manually exits the function.

4.4.2.1AUTO MODE SPANNING

Press Span to enter the span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually. Use the __ arrow keys to toggle between AUTO and MAN span settling. Stop when AUTO appears, blinking, on the display.

Span: Settling: AUTO

<ENT> For Next

Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min

This menu allows the operator to set the time the analyzer should be held in the AUTO span mode, after the readings of the analyzer settle. Five minutes is the default, but it could be adjusted anywhere from 1 to 60 minutes by using the __ keys.

Press Enter to move to the next screen.

Span Val: 000008.00 <ENT>Span <UP>Mod #

Teledyne Analytical Instruments

42

Image 54
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Maintenance OperationAppendix IndexList of Figures Model 3000ZA Front PanelList of Tables Overview Main Features of the AnalyzerIntroduction Typical ApplicationsModel Designations Introduction 3000ZAFront Panel Operator Interface Model 3000ZA Front PanelIntroduction 3000ZA Rear Panel Equipment Interface Model 3000ZA Rear PanelPower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day ZrO2 Sensor Operational TheoryIntroduction Principles of OperationSample System Model 3000ZA Sample SystemElectronics and Signal Processing Zirconium Oxide Sensor Application NotesTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Trace Oxygen Analyzer Installation Rear Panel Connections Gas ConnectionsSample Primary Input Power Electrical ConnectionsEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections Commands via RS-232 Input 2.3 RS-232 PortInstalling the Zirconium Oxide Sensor Testing the SystemTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify System Self-Diagnostic Test LogoutVersion Screen Showing Negative Oxygen Readings Zero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionFixed Range Analysis Signal Output Analyze FunctionRange ID Output Voltage Page Cell Replacement MaintenanceRoutine Maintenance Removing the SensorFuse Replacement Trace Oxygen Analyzer MaintenanceSystem Self Diagnostic Test Installing FusesMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingIf erratic behavior Specifications AppendixTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments