Teledyne 3000ZA operating instructions Fixed Range Analysis

Page 60

Operation

3000ZA

 

 

 

L—10 M—100

H—250 Mode—AUTO

Use the __ arrow keys to blink the range to be set: low (L), medium (M), or high (H).

Use the __ arrow keys to enter the upper value of the range (all ranges begin at 0 ppm). Repeat for each range you want to set. Press Enter to accept the values and return to Analyze mode. (See note below.)

Note: The ranges must be increasing from low to high, for example, if range 1 is set as 0–10 ppm and range 2 is set as 0–100 ppm, range 3 cannot be set as 0– 50 ppm since it is lower than range 2.

4.6.2Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range function by pressing the Range button on the front panel.

Use the __ arrow keys to move the blinking over AUTO.

Use the __ arrow keys to switch from AUTO to FX/LO, FX/MED, or FX/HI to set the instrument on the desired fixed range (low, medium, or high).

L—10

M—100

H—250

Mode—FX/LO

—or—

L—10

M—10 0

H—250

Mode—FX/MED

—or—

L—10

M—10 0

H—250

Mode—FX/HI

Press Escape to re-enter the Analyze mode using the fixed range.

Note: When performing analysis on a fixed range, if the oxygen concentration rises above the upper limit (or default value)

Teledyne Analytical Instruments

48

Image 60
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Maintenance OperationAppendix IndexList of Figures Model 3000ZA Front PanelList of Tables Main Features of the Analyzer IntroductionOverview Typical ApplicationsModel Designations Introduction 3000ZAFront Panel Operator Interface Model 3000ZA Front PanelIntroduction 3000ZA Rear Panel Equipment Interface Model 3000ZA Rear PanelPower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day Operational Theory IntroductionZrO2 Sensor Principles of OperationSample System Model 3000ZA Sample SystemElectronics and Signal Processing Zirconium Oxide Sensor Application NotesTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Installation Unpacking the AnalyzerMounting the Analyzer Trace Oxygen Analyzer Installation Rear Panel Connections Gas ConnectionsSample Primary Input Power Electrical ConnectionsEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections Commands via RS-232 Input 2.3 RS-232 PortInstalling the Zirconium Oxide Sensor Testing the SystemTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify System Self-Diagnostic Test LogoutVersion Screen Showing Negative Oxygen ReadingsZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionFixed Range Analysis Signal Output Analyze FunctionRange ID Output Voltage Page Maintenance Routine MaintenanceCell Replacement Removing the SensorFuse Replacement Trace Oxygen Analyzer MaintenanceSystem Self Diagnostic Test Installing FusesMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingIf erratic behavior Specifications AppendixTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments