Trace Oxygen Analyzer

Operation

 

 

 

Use the __ arrow keys to enter the oxygen-concentration mode. Use the __ arrow keys to blink the digit you are going to modify. Use the __ arrow keys again to change the value of the selected digit. When you have finished typing in the concentration of the span gas you are using, press Enter to begin the Span calibration.

####ppm Span

Slope=#### ppm/s

The beginning span value is shown in the upper left corner of the display. As the span reading settles, the screen displays and updates information on Slope. Spanning automatically ends when the span output corresponds, within tolerance, to the value of the span gas concentration. Then the instrument automatically returns to the analyze mode.

4.4.2.2MANUAL MODE SPANNING

Press Span to start the Span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually.

Span: Settling:MAN

<ENT> For Next

Use the __ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min

This menu allows the operator to set the time the analyzer should be held in the AUTO span mode. It does not have any effect in the MANual mode. Just press Enter key to continue.

Span Val: 000008.00 <ENT>Span <UP>Mod #

Press _ (<UP>) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using. The __ arrows choose the digit, and the __ arrows choose the value of the digit.

Teledyne Analytical Instruments

43

Image 55
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Operation MaintenanceIndex AppendixModel 3000ZA Front Panel List of FiguresList of Tables Typical Applications Main Features of the AnalyzerIntroduction OverviewIntroduction 3000ZA Model DesignationsModel 3000ZA Front Panel Front Panel Operator InterfaceIntroduction 3000ZA Model 3000ZA Rear Panel Rear Panel Equipment InterfacePower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day Principles of Operation Operational TheoryIntroduction ZrO2 SensorModel 3000ZA Sample System Sample SystemZirconium Oxide Sensor Application Notes Electronics and Signal ProcessingTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Unpacking the Analyzer InstallationMounting the Analyzer Trace Oxygen Analyzer Installation Gas Connections Rear Panel ConnectionsSample Electrical Connections Primary Input PowerEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections 2.3 RS-232 Port Commands via RS-232 InputTesting the System Installing the Zirconium Oxide SensorTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify Logout System Self-Diagnostic TestShowing Negative Oxygen Readings Version ScreenZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Analyze Function Signal OutputRange ID Output Voltage Page Removing the Sensor MaintenanceRoutine Maintenance Cell ReplacementTrace Oxygen Analyzer Maintenance Fuse ReplacementInstalling Fuses System Self Diagnostic TestRear Panel Removal Remove only those screws marked with an Major Internal ComponentsTroubleshooting CleaningIf erratic behavior Appendix SpecificationsTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing ListFlow Rate Recommendations Application NotesBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments