8.Are either of the alarms to be defeated?
9.The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms.
10.The defeat function can also be used to reset a latched alarm. (See procedures, below.)
If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions in Section 4.3.3 to enter your password. Once you have clearance to proceed, enter the Alarm function.
Press the Alarm button on the front panel to enter the Alarm function. Make sure that
AL—1 AL—2
Choose Alarm
Set up alarm 1 by moving the blinking over to
Five parameters can be changed on this screen:
• Value of the alarm setpoint,
•
•Defeated?
•Failsafe?
•Latching?
To define the setpoint, use the __ arrow keys to move the blinking over to