Teledyne 3000ZA operating instructions AL-1 AL-2 Choose Alarm

Page 58

Operation

3000ZA

 

 

 

back to non-alarm conditions. This mode requires an alarm to be recognized before it can be reset. In the non-latching mode, the alarm status will terminate when process conditions revert to non-alarm conditions.

8.Are either of the alarms to be defeated?

9.The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms.

10.The defeat function can also be used to reset a latched alarm. (See procedures, below.)

If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions in Section 4.3.3 to enter your password. Once you have clearance to proceed, enter the Alarm function.

Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1is blinking.

AL—1 AL—2

Choose Alarm

Set up alarm 1 by moving the blinking over to AL–1 using the __ arrow keys. Then press Enter to move to the next screen.

AL—1 10 ppm HI

Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

Value of the alarm setpoint, AL–1 #### ppm (oxygen); value can be set from 0 to 10,000.00 ppm although the sensor output is good only to 250 ppm

Out-of-range direction, HI or LO

Defeated? Dft–Y/N(Yes/No)

Failsafe? Fs–Y/N(Yes/No)

Latching? Ltch–Y/N(Yes/No).

To define the setpoint, use the __ arrow keys to move the blinking over to AL–1 ####. Then use the __ arrow keys to change the number. Holding down the key speeds up the incrementing or decrementing. (Remember, the setpoint units are ppm O2.)

Teledyne Analytical Instruments

46

Image 58
Contents Model 3000ZA 3000ZA Specific Model Information Important Notice Safety Messages 3000ZA Vii Table of Contents Maintenance OperationAppendix IndexList of Figures Model 3000ZA Front PanelList of Tables Overview Main Features of the AnalyzerIntroduction Typical ApplicationsModel Designations Introduction 3000ZAFront Panel Operator Interface Model 3000ZA Front PanelIntroduction 3000ZA Rear Panel Equipment Interface Model 3000ZA Rear PanelPower Connection Universal AC power source Scheduled calibrations can vary 2-3 % per day ZrO2 Sensor Operational TheoryIntroduction Principles of OperationSample System Model 3000ZA Sample SystemElectronics and Signal Processing Zirconium Oxide Sensor Application NotesTrace Oxygen Analyzer Operational Theory Component Location Electronic Block Diagram Unpacking the Analyzer InstallationMounting the Analyzer Trace Oxygen Analyzer Installation Rear Panel Connections Gas ConnectionsSample Primary Input Power Electrical ConnectionsEquipment Interface Connector Pin Arrangement Analog Output Connections Alarm Relay Contact Pins Remote Calibration Connections Range ID Relay Connections Remote Probe Connections Commands via RS-232 Input 2.3 RS-232 PortInstalling the Zirconium Oxide Sensor Testing the SystemTrace Oxygen Analyzer Installation Operation Using the Data Entry and Function Buttons Hierarchy of Available Functions System Function TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection Pswd Restrictions Removed Retype PWD To Verify System Self-Diagnostic Test LogoutVersion Screen Showing Negative Oxygen ReadingsZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal #### ppm Span Slope=#### ppm/s Span Failure Alarms Function AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionFixed Range Analysis Signal Output Analyze FunctionRange ID Output Voltage Page Cell Replacement MaintenanceRoutine Maintenance Removing the SensorFuse Replacement Trace Oxygen Analyzer MaintenanceSystem Self Diagnostic Test Installing FusesMajor Internal Components Rear Panel Removal Remove only those screws marked with anCleaning TroubleshootingIf erratic behavior Specifications AppendixTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel MountApplication Notes Flow Rate RecommendationsBY-PASS Conversions PSI Scfh = 0.476 Slpm Index Trace Oxygen Analyzer Index Relay connections, 22 relay contacts Trace Oxygen Analyzer Index Teledyne Analytical Instruments